Repeatability and Reproducibility of Foveal Avascular Zone Area Measurement on Normal Eyes by Different Optical Coherence Tomography Angiography Instruments.
Ophthalmic Research 2018
PURPOSE: To compare the foveal avascular zone (FAZ) area measurements produced by different optical coherence tomography angiography (OCTA).
METHODS: Healthy enrolled volunteers underwent OCTA using 2 different devices: Spectralis HRA+OCTA (Heidelberg Engineering, Heidelberg, Germany) and RS-3000 Advance (Nidek, Gamagori, Japan). Two graders measured FAZ in both superficial (SCP) and deep (DCP) retinal capillary plexuses. The SCP and DCP en face images were visualized automatically segmenting 2 separate slabs defined by the arbitrary segmentation lines created by the software of each OCT device. One grader repeated each measure twice.
RESULTS: Fifty-nine eyes were included. The mean FAZ was 0.33 ± 0.09 mm2 at the SCP and 0.57 ± 0.17 mm2 at the DCP measured with RS-3000 versus 0.30 ± 0.08 and 0.35 ± 0.08 mm2, respectively, measured with Spectralis. The measurements of the 2 devices were significantly different (p < 0.0001). The intraoperator agreement was excellent at the SCP (intraclass correlation coefficient, ICC: 0.97 with Spectralis and 0.96 with RS-3000). At the DCP, it was good with Spectralis and fair with RS-3000 (ICC: 0.85 and 0.64, respectively). The interoperator agreement was excellent for Spectralis and good for RS-3000 at the SCP (ICC: 0.97 and 0.93, respectively). It was good at the DCP with both devices (ICC: 0.74 with RS-3000 and 0.81 with Spectralis).
CONCLUSIONS: FAZ measurements obtained with different OCTA devices differ. These findings should be considered in follow-up studies of patients with retinal vascular diseases.
METHODS: Healthy enrolled volunteers underwent OCTA using 2 different devices: Spectralis HRA+OCTA (Heidelberg Engineering, Heidelberg, Germany) and RS-3000 Advance (Nidek, Gamagori, Japan). Two graders measured FAZ in both superficial (SCP) and deep (DCP) retinal capillary plexuses. The SCP and DCP en face images were visualized automatically segmenting 2 separate slabs defined by the arbitrary segmentation lines created by the software of each OCT device. One grader repeated each measure twice.
RESULTS: Fifty-nine eyes were included. The mean FAZ was 0.33 ± 0.09 mm2 at the SCP and 0.57 ± 0.17 mm2 at the DCP measured with RS-3000 versus 0.30 ± 0.08 and 0.35 ± 0.08 mm2, respectively, measured with Spectralis. The measurements of the 2 devices were significantly different (p < 0.0001). The intraoperator agreement was excellent at the SCP (intraclass correlation coefficient, ICC: 0.97 with Spectralis and 0.96 with RS-3000). At the DCP, it was good with Spectralis and fair with RS-3000 (ICC: 0.85 and 0.64, respectively). The interoperator agreement was excellent for Spectralis and good for RS-3000 at the SCP (ICC: 0.97 and 0.93, respectively). It was good at the DCP with both devices (ICC: 0.74 with RS-3000 and 0.81 with Spectralis).
CONCLUSIONS: FAZ measurements obtained with different OCTA devices differ. These findings should be considered in follow-up studies of patients with retinal vascular diseases.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app