Techniques for trans-catheter retrieval of embolized Nit-Occlud ® PDA-R and ASD-R devices.
Catheterization and Cardiovascular Interventions 2018 Februrary 16
BACKGROUND: Nit-Occlud® (atrial septal defect) ASD-R and (patent ductus arteriosus) PDA-R devices are used outside the United States for percutaneous closure of the patent ductus arteriosus and atrial septal defects. When embolization occurs, these devices have been difficult to retrieve.
METHODS: Bench simulations of retrieval of PDA-R and ASD-R devices were performed in a vascular model. Retrieval of each device was attempted using snare techniques or with bioptome forceps with a range of devices. The same devices were then intentionally embolized in an animal model. Retrieval methods were systematically tested in a range of sheath sizes, and graded in terms of difficulty and retrieval time.
RESULTS: Devices that were grasped by the bioptome in the center of the proximal part of the devices were easily retrieved in both models. Bench studies determined the minimum sheath sizes needed for retrieval of each device with this method. In general sheathes two french sizes greater than the delivery sheath were successful with this technique. Three out of the four PDA-R devices were successfully retrieved in vivo. Two were retrieved by grasping the middle of the PA end of the PDA-R device with a Maslanka bioptome and one small PDA-R device was retrieved using a 10 mm Snare. Four of the five ASD-R devices were retrieved successfully grasping the right atrial ASD-R disc or by passing a wire through the device and snaring this loop. For ASD-R 28 and 30 mm devices, a double bioptome technique was needed to retrieve the device.
CONCLUSION: ASD-R and PDA-R devices can be successfully retrieved in the catheterization lab. It is critical to grab the center portion of the right atrial disc of the ASD-R device or pulmonary portion of the PDA-R device and to use adequately sized sheathes.
METHODS: Bench simulations of retrieval of PDA-R and ASD-R devices were performed in a vascular model. Retrieval of each device was attempted using snare techniques or with bioptome forceps with a range of devices. The same devices were then intentionally embolized in an animal model. Retrieval methods were systematically tested in a range of sheath sizes, and graded in terms of difficulty and retrieval time.
RESULTS: Devices that were grasped by the bioptome in the center of the proximal part of the devices were easily retrieved in both models. Bench studies determined the minimum sheath sizes needed for retrieval of each device with this method. In general sheathes two french sizes greater than the delivery sheath were successful with this technique. Three out of the four PDA-R devices were successfully retrieved in vivo. Two were retrieved by grasping the middle of the PA end of the PDA-R device with a Maslanka bioptome and one small PDA-R device was retrieved using a 10 mm Snare. Four of the five ASD-R devices were retrieved successfully grasping the right atrial ASD-R disc or by passing a wire through the device and snaring this loop. For ASD-R 28 and 30 mm devices, a double bioptome technique was needed to retrieve the device.
CONCLUSION: ASD-R and PDA-R devices can be successfully retrieved in the catheterization lab. It is critical to grab the center portion of the right atrial disc of the ASD-R device or pulmonary portion of the PDA-R device and to use adequately sized sheathes.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app