We have located links that may give you full text access.
Diagnostic reference levels for common computed tomography (CT) examinations: results from the first Nigerian nationwide dose survey.
Journal of Radiological Protection : Official Journal of the Society for Radiological Protection 2018 June
PURPOSE: To explore doses from common adult computed tomography (CT) examinations and propose national diagnostic reference levels (nDRLs) for Nigeria.
MATERIALS AND METHODS: This retrospective study was approved by the Nnamdi Azikiwe University and University Teaching Hospital Institutional Review Boards (IRB: NAUTH/CS/66/Vol8/84) and involved dose surveys of adult CT examinations across the six geographical regions of Nigeria and Abuja from January 2016 to August 2017. Dose data of adult head, chest and abdomen/pelvis CT examinations were extracted from patient folders. The median, 75th and 25th percentile CT dose index volume (CTDIvol ) and dose-length-product (DLP) were computed for each of these procedures. Effective doses (E) for these examinations were estimated using the k conversion factor as described in the ICRP publication 103 (EDLP = k × DLP ).
RESULTS: The proposed 75th percentile CTDIvol for head, chest, and abdomen/pelvis are 61 mGy, 17 mGy, and 20 mGy, respectively. The corresponding DLPs are 1310 mGy.cm, 735 mGy.cm, and 1486 mGy.cm respectively. The effective doses were 2.75 mSv (head), 10.29 mSv (chest), and 22.29 mSv (abdomen/pelvis).
CONCLUSION: Findings demonstrate wide dose variations within and across centres in Nigeria. The results also show CTDIvol comparable to international standards, but considerably higher DLP and effective doses.
MATERIALS AND METHODS: This retrospective study was approved by the Nnamdi Azikiwe University and University Teaching Hospital Institutional Review Boards (IRB: NAUTH/CS/66/Vol8/84) and involved dose surveys of adult CT examinations across the six geographical regions of Nigeria and Abuja from January 2016 to August 2017. Dose data of adult head, chest and abdomen/pelvis CT examinations were extracted from patient folders. The median, 75th and 25th percentile CT dose index volume (CTDIvol ) and dose-length-product (DLP) were computed for each of these procedures. Effective doses (E) for these examinations were estimated using the k conversion factor as described in the ICRP publication 103 (EDLP = k × DLP ).
RESULTS: The proposed 75th percentile CTDIvol for head, chest, and abdomen/pelvis are 61 mGy, 17 mGy, and 20 mGy, respectively. The corresponding DLPs are 1310 mGy.cm, 735 mGy.cm, and 1486 mGy.cm respectively. The effective doses were 2.75 mSv (head), 10.29 mSv (chest), and 22.29 mSv (abdomen/pelvis).
CONCLUSION: Findings demonstrate wide dose variations within and across centres in Nigeria. The results also show CTDIvol comparable to international standards, but considerably higher DLP and effective doses.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app