Evaluation of Hematocrit Influence on Self-Monitoring of Blood Glucose Based on ISO 15197:2013: Comparison of a Novel System With Five Systems With Different Hematocrit Ranges.
Journal of Diabetes Science and Technology 2018 March
INTRODUCTION: ISO 15197:2013 recommends testing procedures and acceptance criteria for the evaluation of influence quantities such as hematocrit on measurement results with systems for self-monitoring of blood glucose (SMBG). In this study, hematocrit influence was evaluated for a novel SMBG system (system A) and five other systems with different hematocrit ranges based on ISO 15197:2013.
METHODS: Test procedures were performed with one test strip lot for each system. Each system was tested within the hematocrit range indicated in the manufacturer's labeling (system A: 10-65%, B: 15-65%, C: 20-60%, D: 35-60%, E: 30-60%, F: 30-55%). According to ISO 15197:2013, clause 6.4.2, venous blood samples were used for the evaluation of hematocrit influence. The evaluation was performed for three glucose concentration categories (30-50 mg/dL, 96-144 mg/dL, and 280-420 mg/dL). For each glucose concentration category, at least five different hematocrit levels were investigated.
RESULTS: The novel system A and systems B, E, and F complied with the tested lot with the defined criteria and showed ≤10 mg/dL and ≤10% difference between the test sample and the respective control sample with a hematocrit value of 42% ± 2% for BG concentrations <100 mg/dL and ≥100 mg/dL, respectively. Two systems showed >10% difference at glucose concentrations ≥100 mg/dL.
CONCLUSIONS: Remarkable hematocrit influence within the labeled hematocrit range was obtained in two systems with the tested reagent system lot. Adequate SMBG systems should be carefully chosen by patients and their health care professionals, particularly for patients with increased and decreased hematocrit values.
METHODS: Test procedures were performed with one test strip lot for each system. Each system was tested within the hematocrit range indicated in the manufacturer's labeling (system A: 10-65%, B: 15-65%, C: 20-60%, D: 35-60%, E: 30-60%, F: 30-55%). According to ISO 15197:2013, clause 6.4.2, venous blood samples were used for the evaluation of hematocrit influence. The evaluation was performed for three glucose concentration categories (30-50 mg/dL, 96-144 mg/dL, and 280-420 mg/dL). For each glucose concentration category, at least five different hematocrit levels were investigated.
RESULTS: The novel system A and systems B, E, and F complied with the tested lot with the defined criteria and showed ≤10 mg/dL and ≤10% difference between the test sample and the respective control sample with a hematocrit value of 42% ± 2% for BG concentrations <100 mg/dL and ≥100 mg/dL, respectively. Two systems showed >10% difference at glucose concentrations ≥100 mg/dL.
CONCLUSIONS: Remarkable hematocrit influence within the labeled hematocrit range was obtained in two systems with the tested reagent system lot. Adequate SMBG systems should be carefully chosen by patients and their health care professionals, particularly for patients with increased and decreased hematocrit values.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app