Precision of the Connection Between Implant and Standard or Computer-Aided Design/Computer-Aided Manufacturing Abutments: A Novel Evaluation Method.

PURPOSE: The aim of this in vitro study was to verify whether or not stock and computer-aided design/computer-aided manufacturing (CAD/CAM) abutments show similar precision in the connection with the respective implants.

MATERIALS AND METHODS: Ten CAD/CAM titanium abutments were compared with 10 stock titanium abutments. Each abutment fit a regular-platform implant (Institute Straumann). Implants and abutments were measured independently and then connected. During the connection procedure, the torque was measured using a six-axes load cell. Then, outer geometric features of the implant-abutment connection were measured again. Finally, the assembly was sectioned to provide the analysis of inner surfaces in contact. The geometric measurements were performed using a multisensored opto-mechanical coordinate measuring machine. The following parameters were measured and compared for the CAD/CAM and stock titanium abutment groups, respectively: width of interference and interference length between the conical surfaces of the implant and abutment; and volume of material involved in the implant-abutment connection.

RESULTS: Interference width mean ± SD values of 18 ± 0.5 and 14 μ 0.5 μm were calculated for the stock and CAD/CAM titanium abutment groups, respectively. The difference was statistically significant (P = .02). Furthermore, the interference length mean ± SD values of 763 ± 10 and 816 ± 43 μm were calculated for stock and CAD/CAM titanium abutment groups, respectively. The difference was also statistically significant (P = .04). Finally, the volume of material involved in the implant-abutment connection was compared between stock and CAD/CAM titanium abutment groups; the mean ± SD values of 0.134 ± 0.014 and 0.108 ± 0.023 mm3 were significantly different (P = .009).

CONCLUSION: Both standard and CAD/CAM abutment groups showed a three-dimensional (3D) seal activation after the screw tightening. Nevertheless, stock titanium abutments showed a significantly higher volume of material involved in the implant-abutment connection compared with that of CAD/CAM titanium abutments.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app