Optimizing reverse shoulder arthroplasty component position in the setting of advanced arthritis with posterior glenoid erosion: a computer-enhanced range of motion analysis.

BACKGROUND: Our study purpose was to determine the optimal glenoid and humeral reverse shoulder arthroplasty (RSA) component design and position in osteoarthritic shoulders with severe glenoid retroversion deformities.

METHODS: Computed tomography scans from 10 subjects were analyzed with advanced software including RSA range of motion (ROM) analysis. Variables included glenoid component retroversion of 0°, 5°, 10°, 15°, and 20° and baseplate lateralization of 0, 5, and 10 mm. Humeral variables included 135°, 145°, and 155° angle of inclination (AOI) combined with variable humeral offset.

RESULTS: Glenoid component lateralization had the greatest influence on ROM. In comparing each ROM direction among all lateralization options independently, there were significantly greater adduction, abduction, external rotation, extension, and flexion motions with progressively greater lateralization. Internal rotation motion was greater at 10 mm only. In analyzing the effects of glenoid version independently, no differences in adduction or abduction ROM were seen. With greater retroversion, decreased external rotation and extension motion was noted; however, greater internal rotation and flexion motion was seen with the exception of flexion at 10 mm of lateralization. For adduction, external rotation, and extension, a more valgus AOI resulted in less ROM at each progressively greater AOI independent of humeral lateralization. Internal rotation and flexion motions were greater with a more varus AOI but not significant between each inclination angle. Abduction ROM was maximized with a more valgus AOI. Humeral lateralization had no effect on ROM.

CONCLUSIONS: In the setting of RSA for advanced glenoid osteoarthritic deformities, optimal ROM is achieved with 10-mm baseplate lateralization and neutral to 5° of retroversion mated to a humeral implant with a varus (135°) inclination angle.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app