LumenRECON Guidewire: Pilot Study of a Novel, Nonimaging Technology for Accurate Vessel Sizing and Delivery of Therapy in Femoropopliteal Disease.

BACKGROUND: Proper vessel sizing during endovascular interventions is crucial to avoid adverse procedural and clinical outcomes. LumenRECON (LR) is a novel, nonimaging, 0.035-inch wire-based technology that uses the physics-based principle of Ohm's law to provide a simple, real-time luminal size while also providing a platform for therapy delivery. This study evaluated the accuracy, reliability, and safety of the LR system in patients presenting for a femoropopliteal artery intervention.

METHODS AND RESULTS: This multicenter, prospective pilot study of 24 patients presenting for peripheral intervention compared LR measurements of femoropopliteal artery size to angiographic visual estimation, duplex ultrasound, quantitative angiography, and intravascular ultrasound. The primary effectiveness and safety end point was comparison against core laboratory adjudicated intravascular ultrasound values and major adverse events, respectively. Additional preclinical studies were also performed in vitro and in vivo in swine to determine the accuracy of the LR guidewire system. No intra- or postprocedure device-related adverse events occurred. A balloon or stent was successfully delivered in 12 patients (50%) over the LR wire. Differences in repeatability between successive LR measurements was 2.5±0.40% ( R 2 =0.96) with no significant bias. Differences in measurements of LR to other modalities were 0.5±1.7%, 5.0±1.8%, -1.5±2.0%, and 6.8±3.4% for intravascular ultrasound core laboratory, quantitative angiography, angiographic, and duplex ultrasound, respectively.

CONCLUSIONS: This study demonstrates that through a physics-based principle, LR provides a real-time, safe, reproducible, and accurate vessel size of the femoropopliteal artery during intervention and can additionally serve as a conduit for therapy delivery over its wire-based platform.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app