Quantitative Assessment of Inflow Malposition in Two Continuous-Flow Left Ventricular Assist Devices.
Annals of Thoracic Surgery 2018 May
BACKGROUND: We previously investigated preoperative variables associated with qualitative inflow cannula malposition in the HeartMate II (Thoratec-Abbott, Abbott Park, IL) continuous-flow left ventricular assist device. In this report, we assess inflow cannula malposition quantitatively in recipients of both the HeartMate II and the HeartWare (Medtronic-HeartWare, Minneapolis, MN) and examine its association with device thrombosis.
METHODS: Malposition was quantified based on angular deviation from a hypothetic ideal inflow cannula position in two orthogonal computed tomography imaging planes. Ideal position lies on a line from the apex to the center of the mitral valve. Positive anterior plane angulation indicates deviation toward the superior free wall; negative, toward the inferior wall. Positive lateral plane angulation indicates deviation toward the septum; negative, toward the lateral wall. Device thrombosis was assessed based on clinical criteria.
RESULTS: Fifty-four HeartMate II patients and 68 HeartWare patients were analyzed. Inflow cannula deviation was significantly higher for HeartMate II than for HeartWare (anterior plane angle 36.7 ± 16.8 versus -18.7 ± 11.6 degrees, p < 0.001; lateral plane angle 23.7 ± 20.1 versus 0.2 ± 15.0 degrees, p < 0.001. Pump thrombosis occurred in 31% of HeartMate II patients and 2.9% of HeartWare patients (p < 0.001). In a multivariate model, HeartMate II and increasing inflow cannula deviation toward the septum were associated with higher thrombosis risk (odds ratio 1.35 per 10-degree increase).
CONCLUSIONS: We found distinct device-dependent differences in inflow cannula positioning and thrombosis, with HeartWare showing both less malposition and less thrombosis. Malposition toward the ventricular septum may contribute to pump thrombosis through a vicious cycle of suction events, low flow, and speed reduction.
METHODS: Malposition was quantified based on angular deviation from a hypothetic ideal inflow cannula position in two orthogonal computed tomography imaging planes. Ideal position lies on a line from the apex to the center of the mitral valve. Positive anterior plane angulation indicates deviation toward the superior free wall; negative, toward the inferior wall. Positive lateral plane angulation indicates deviation toward the septum; negative, toward the lateral wall. Device thrombosis was assessed based on clinical criteria.
RESULTS: Fifty-four HeartMate II patients and 68 HeartWare patients were analyzed. Inflow cannula deviation was significantly higher for HeartMate II than for HeartWare (anterior plane angle 36.7 ± 16.8 versus -18.7 ± 11.6 degrees, p < 0.001; lateral plane angle 23.7 ± 20.1 versus 0.2 ± 15.0 degrees, p < 0.001. Pump thrombosis occurred in 31% of HeartMate II patients and 2.9% of HeartWare patients (p < 0.001). In a multivariate model, HeartMate II and increasing inflow cannula deviation toward the septum were associated with higher thrombosis risk (odds ratio 1.35 per 10-degree increase).
CONCLUSIONS: We found distinct device-dependent differences in inflow cannula positioning and thrombosis, with HeartWare showing both less malposition and less thrombosis. Malposition toward the ventricular septum may contribute to pump thrombosis through a vicious cycle of suction events, low flow, and speed reduction.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app