Goldmann and error correcting tonometry prisms compared to intracameral pressure.

BMC Ophthalmology 2018 January 5
BACKGROUND: Compare Goldmann applanation tonometer (GAT) prism and correcting applanation tonometry surface (CATS) prism to intracameral intraocular pressure (IOP), in vivo and in vitro.

METHODS: Pressure transducer intracameral IOP was measured on fifty-eight (58) eyes undergoing cataract surgery and the IOP was modulated manometrically to 10, 20, and 40 mmHg. Simultaneously, IOP was measured using a Perkins tonometer with a standard GAT prism and a CATS prism at each of the intracameral pressures. Statistical comparison was made between true intracameral pressures and the two prism measurements. Differences between the two prism measurements were correlated to central corneal thickness (CCT) and corneal resistance factor (CRF). Human cadaver eyes were used to assess measurement repeatability.

RESULTS: The CATS tonometer prism measured closer to true intracameral IOP than the GAT prism by 1.7+/-2.7 mmHg across all pressures and corneal properties. The difference in CATS and GAT measurements was greater in thin CCT corneas (2.7+/-1.9 mmHg) and low resistance (CRF) corneas (2.8+/-2.1 mmHg). The difference in prisms was negligible at high CCT and CRF values. No difference was seen in measurement repeatability between the two prisms.

CONCLUSION: A CATS prism in Goldmann tonometer armatures significantly improve the accuracy of IOP measurement compared to true intracameral pressure across a physiologic range of IOP values. The CATS prism is significantly more accurate compared to the GAT prism in thin and less rigid corneas. The in vivo intracameral study validates mathematical models and clinical findings in IOP measurement between the GAT and CATS prisms.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app