A Study to Determine Whether the Volume-Weighted Computed Tomography Dose Index Gives Reasonable Estimates of Organ Doses for Thai Patients Undergoing Abdomen and Pelvis Computed Tomography Examinations.

Introduction: Values for the CTDIvol , which is displayed on scanner consoles, give doses relative to a phantom much larger than most Thai patients, and the CTDIvol does not take account of differences in patient size, which affect organ doses.

Objective: The purpose of this study was to evaluate relationships for size specific dose estimate (SSDE) and volume weighted computed tomography (CT) dose index (CTDIvol ) with patient size for CT scanners operating under automatic tube current modulation (ATCM).

Methods: Retrospective data from 244 patients who had undergone abdomen and pelvis examination on GE and Siemens CT scanners were included in this study. The combination of anteroposterior (AP) and lateral dimensions at the level of the first lumbar vertebra (L1) was used to represent patient size. Image noise within the liver was measured, and values of the absorbed dose for organs covered by the primary beam such as the liver, stomach and kidney were calculated using methods described in the literature. Values of CTDIvol were recorded and SSDE calculated according to the American Association of Physics in Medicine (AAPM) Report No.204. Linear regression models were used to evaluate the relationship between SSDE, CTDIvol , image noise and patient size.

Results: SSDE is 20%-50% larger than the CTDIvol , with values for larger patients being more representative. Both the CTDIvol and image noise decreased with patient size for Siemens scanners, but the decline in SSDE was less significant. For the GE scanner, the CTDIvol was a factor of 3-4 lower in small patients compared to larger ones, while the SSDE only decreased by a factor of two. Noise actually decreased slightly with patient size.

Conclusion: Values of SSDE were similar to the doses calculated for the liver, stomach and kidney, which are covered by the primary beam, confirming that it provides a good estimate of organ-absorbed dose.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app