Establishing the appropriate position of proximal humerus locking plates: a cadaveric study of five plating systems and their anatomic reference measurements.
Shoulder & Elbow 2018 January
Background: The present study aimed to determine anatomic references for the placement of five proximal humerus locking plates.
Methods: Five proximal humerus locking-plate systems were placed on six human shoulder cadavers. Plates were positioned by fluoroscopic confirmation so that the inferior oblique screw was within 5 mm of the inferomedial cortex. Plate position was measured using the superior border of the pectoralis major tendon (PMT) to the bottom of the first slotted or nonlocking hole and top of the plate to the top of the greater tuberosity. The distance from the PMT insertion to the top of the humeral head was measured as a control.
Results: There was consistency within each plating system for both the distance from the PMT insertion to the first hole and the top of the plate to the greater tuberosity: Synthes first-generation [mean (SD) 13.7 mm (3.1 mm); 10 mm (1.3) mm], Synthes second-generation [28.2 mm (2.2 mm); 18.5 mm (2.7 mm)], Biomet OptiLock® [25.5 mm (2.7 mm); 18.7 mm (2 mm)], Stryker AxSOS® [5 mm (2.8 mm); 12.3 mm (3.3 mm)] and Acumed Polarus® [9.5 mm (1.8 mm); 14.8 mm (1.6 mm)].
Conclusions: The present study provides measurements that improve the accuracy of plate positioning for five plating systems.
Methods: Five proximal humerus locking-plate systems were placed on six human shoulder cadavers. Plates were positioned by fluoroscopic confirmation so that the inferior oblique screw was within 5 mm of the inferomedial cortex. Plate position was measured using the superior border of the pectoralis major tendon (PMT) to the bottom of the first slotted or nonlocking hole and top of the plate to the top of the greater tuberosity. The distance from the PMT insertion to the top of the humeral head was measured as a control.
Results: There was consistency within each plating system for both the distance from the PMT insertion to the first hole and the top of the plate to the greater tuberosity: Synthes first-generation [mean (SD) 13.7 mm (3.1 mm); 10 mm (1.3) mm], Synthes second-generation [28.2 mm (2.2 mm); 18.5 mm (2.7 mm)], Biomet OptiLock® [25.5 mm (2.7 mm); 18.7 mm (2 mm)], Stryker AxSOS® [5 mm (2.8 mm); 12.3 mm (3.3 mm)] and Acumed Polarus® [9.5 mm (1.8 mm); 14.8 mm (1.6 mm)].
Conclusions: The present study provides measurements that improve the accuracy of plate positioning for five plating systems.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app