We have located links that may give you full text access.
Clinical Trial
Comparative Study
Journal Article
Comparative evaluation of neonatal bilirubin using Radiometer whole blood co-oximetry and plasma bilirubin methods from Roche Diagnostics and Ortho Clinical Diagnostics.
Clinical Biochemistry 2018 March
BACKGROUND: Clinically significant variation has been reported within and between plasma and whole blood total bilirubin methods used to identify neonates for whom clinical intervention for hyperbilirubinemia may be required.
OBJECTIVE: To evaluate total bilirubin measurements between the Radiometer whole blood co-oximeter and plasma bilirubin methods from Roche Diagnostics and Ortho Clinical Diagnostics using neonatal specimens.
METHODS: Total bilirubin levels were analyzed by whole blood co-oximetry (Radiometer® ABL90). Specimens were centrifuged and plasma analyzed for total bilirubin with a diazo method (Roche Cobas® C-601) and a reflectance spectrophotometric BuBc dry film method (Ortho Clinical Diagnostics VITROS® 350). Results were evaluated by regression, Bland-Altman comparisons and t-tests.
RESULTS: The patient correlation study yielded the following regression equations in μmol/L: a) Radiometer=1.03 Roche - 3.5μmol/L b) Radiometer=0.98 Ortho - 5.7μmol/L c) Roche=0.97 Ortho - 2.4μmol/L. The mean bias over the range of total bilirubin levels examined was -1.0μmol/L for the Radiometer versus the Roche (p≤0.305); -4.4μmol/L for the Radiometer versus Ortho (p≤0.005) and -4.4μmol/L for the Roche versus Ortho (p≤0.002).
CONCLUSIONS: Whole blood total bilirubin measurement using the Radiometer ABL90 blood gas analyzer provides accurate and precise results compared to the Roche plasma diazo method. Compared to the reflectance spectrophotometric method, results are precise and had a small but statistically significant bias of -4.4μmol/L.
OBJECTIVE: To evaluate total bilirubin measurements between the Radiometer whole blood co-oximeter and plasma bilirubin methods from Roche Diagnostics and Ortho Clinical Diagnostics using neonatal specimens.
METHODS: Total bilirubin levels were analyzed by whole blood co-oximetry (Radiometer® ABL90). Specimens were centrifuged and plasma analyzed for total bilirubin with a diazo method (Roche Cobas® C-601) and a reflectance spectrophotometric BuBc dry film method (Ortho Clinical Diagnostics VITROS® 350). Results were evaluated by regression, Bland-Altman comparisons and t-tests.
RESULTS: The patient correlation study yielded the following regression equations in μmol/L: a) Radiometer=1.03 Roche - 3.5μmol/L b) Radiometer=0.98 Ortho - 5.7μmol/L c) Roche=0.97 Ortho - 2.4μmol/L. The mean bias over the range of total bilirubin levels examined was -1.0μmol/L for the Radiometer versus the Roche (p≤0.305); -4.4μmol/L for the Radiometer versus Ortho (p≤0.005) and -4.4μmol/L for the Roche versus Ortho (p≤0.002).
CONCLUSIONS: Whole blood total bilirubin measurement using the Radiometer ABL90 blood gas analyzer provides accurate and precise results compared to the Roche plasma diazo method. Compared to the reflectance spectrophotometric method, results are precise and had a small but statistically significant bias of -4.4μmol/L.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app