We have located links that may give you full text access.
A comparative evaluation of accuracy of three electronic apex locators using histological section as gold standard: An ex vivo study.
Journal of Conservative Dentistry : JCD 2017 July
Introduction: The aim of the present study was to comparatively evaluate the accuracy of iRoot, iPex II, and Propex pixi apex locator using histological sections as the gold standard.
Materials and Methods: Thirty patients indicated for extraction of single-rooted permanent teeth with single canal system were selected. Working lengths (WLs) of teeth were determined using iRoot, iPex II, and Propex pixi. Teeth were then extracted, and the files were reintroduced to the anatomic apex to measure anatomic canal length (ACL) and fixed at the ACL using flowable composite. The apical 4 mm of the roots were longitudinally shaved away to visualize the canal under a stereomicroscope at ×24 magnification. Digital photographs were evaluated to measure the distance between the major diameter and minor diameter. Thus, the WL, that is, the minor diameter length (MDL) was ascertained.
Results: Measurements of mean WLs within ±0.5 mm of minor diameter were 90% acceptable for iRoot, 86.66% for iPex II, and 80% for Propex pixi when compared with mean MDL as obtained from the histological sections.
Conclusions: All apex locators have been shown to produce acceptable level of accuracy which clearly indicates their reliability in determining the WL.
Materials and Methods: Thirty patients indicated for extraction of single-rooted permanent teeth with single canal system were selected. Working lengths (WLs) of teeth were determined using iRoot, iPex II, and Propex pixi. Teeth were then extracted, and the files were reintroduced to the anatomic apex to measure anatomic canal length (ACL) and fixed at the ACL using flowable composite. The apical 4 mm of the roots were longitudinally shaved away to visualize the canal under a stereomicroscope at ×24 magnification. Digital photographs were evaluated to measure the distance between the major diameter and minor diameter. Thus, the WL, that is, the minor diameter length (MDL) was ascertained.
Results: Measurements of mean WLs within ±0.5 mm of minor diameter were 90% acceptable for iRoot, 86.66% for iPex II, and 80% for Propex pixi when compared with mean MDL as obtained from the histological sections.
Conclusions: All apex locators have been shown to produce acceptable level of accuracy which clearly indicates their reliability in determining the WL.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app