Noninvasive Assessment of Intra-Abdominal Pressure Using Ultrasound-Guided Tonometry: A Proof-of-Concept Study.

Shock 2018 December
BACKGROUND: Intra-abdominal hypertension jeopardizes abdominal organ perfusion and venous return. Contemporary recognition of elevated intra-abdominal pressure (IAP) plays a crucial role in reducing mortality and morbidity. We evaluated ultrasound-guided tonometry in this context hypothesizing that the vertical chamber diameter of this device inversely correlates with IAP.

METHODS: IAP was increased in six 5 mmHg steps to 40 mmHg by instillation of normal saline into the peritoneal cavity of eight anesthetized pigs. Liver and renal blood flows (ultrasound transit time), intravesical, intraperitoneal, and end-inspiratory plateau pressures were recorded. For ultrasound-based assessment of IAP (ultrasound-guided tonometry), a pressure-transducing, compressible chamber was fixed at the tip of a linear ultrasound probe, and the system was applied on the abdominal wall using different predetermined levels of external pressure. At each IAP level (reference: intravesical pressure), two investigators measured the vertical diameter of this chamber.

RESULTS: All abdominal flows decreased (by 39%-58%), and end-inspiratory plateau pressure increased from 15 mbar (14-17 mbar) to 38 mbar (33-42 mbar) (median, range) with increasing IAP (all P < 0.01). Vertical chamber diameter decreased from 14.9 (14.6-15.2) mm to12.8 (12.4-13.4) mm with increasing IAP. Coefficients of variations between and within observers regarding change of the vertical tonometry chamber diameter were small (all <4%), and the results were independent of the externally applied pressure level on the ultrasound probe. Correlation of IAP and vertical pressure chamber distance was highly significant (r = -1, P = 0.0004). Ultrasound-guided tonometry could discriminate between normal (baseline) pressure and 15 mmHg, between 15 and 25 mmHg) and between 25 and 40 mmHg IAP (all P ≤ 0.18). Similar results were obtained for end-inspiratory plateau pressures.

CONCLUSIONS: In our model, values obtained by ultrasound-guided tonometry correlated significantly with IAPs. The method was able to discriminate between normal, moderately, and markedly increased IAP values.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app