Patency and in vivo compatibility of bacterial nanocellulose grafts as small-diameter vascular substitute.

OBJECTIVE: Despite the clinical success of large-diameter vascular grafts, synthetic grafts in small-diameter vessels are of limited use because of their poor patency rates. Previous experiments of our group provided evidence for good biocompatibility of bacterial nanocellulose (BNC) as a small-vessel graft in the carotid artery in sheep. However, the patency rate of our first-generation tubes after 3 months was only 50%. To advance our concept, we now used modified second-generation tubes with diminished wall thickness and a smoother inner surface to reduce the thrombogenic potential. The aim was to investigate mechanical characteristics of modified second-generation BNC tubes, to evaluate in vivo performance and biocompatibility, and to analyze patency rates.

METHODS: We replaced the right carotid artery of 23 sheep with second-generation BNC tubes. Compared with our first-generation tubes, tubes were modified with different surface properties and diminished wall thickness (inner diameter, 4.0-5.0 mm; wall thickness, 1.0-2.5 mm; length, 100 mm) to generate a smoother inner surface with reduced thrombogenic potential and a more porous outer zone, allowing easier cell immigration.

RESULTS: At the end of the investigational period, BNC tubes were explanted and grafts were processed for histopathologic analysis. Histologic analysis revealed no acute signs of foreign body reaction such as immigration of giant cells or other acute inflammatory reaction and therefore provided evidence for good biocompatibility of the second-generation tubes. However, all grafts of the sheep without antiplatelet therapy were occluded after 9 months, whereas grafts in sheep receiving dual platelet inhibition showed a patency rate of 67% (six of nine grafts). Further modified grafts revealed a patency rate of 80% (four of five grafts remained open).

CONCLUSIONS: Patency rates of the second-generation tubes could be substantially improved compared with our first-generation tubes. However, poor patency rates of tissue-engineered blood vessels still limit their use in clinical studies. Further efforts in terms of in vitro and in vivo studies are essential to improve grafts of BNC.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app