In vitro performance of prefilled CO 2 absorbers with the Zeus®.

Low fresh gas flows (FGFs) decrease the use of anesthetic gases, but increase CO2 absorbent usage. CO2 absorbent usage remains poorly quantified. The goal of this study is to determine canister life of 8 commercially available CO2 absorbent prepacks with the Zeus® . Pre-packed CO2 canisters of 8 different brands were tested in vitro: Amsorb Plus, Spherasorb, LoFloSorb, LithoLyme, SpiraLith, SpheraSorb, Drägersorb 800+, Drägersorb Free, and CO2ntrol. CO2 (160 mL min- 1 ) flowed into the tip of a 2 L breathing bag that was ventilated with a tidal volume of 500 mL, a respiratory rate of 10/min, and an I:E ratio of 1:1 using the controlled mechanical ventilation mode of the Zeus® (Dräger, Lubeck, Germany). In part I, canister life of 5 canisters each of 2 different lots of each brand was determined with a 350 mL min- 1 FGF. Canister life is the time it takes for the inspired CO2 concentration (FI CO2 ) to rise to 0.5%. In part II, canister life was measured accross a FGF range of 0.25 to 4 L min- 1 for Drägersorb 800+ (2 lots) and SpiraLith (1 lot). In part III, the calculated canister life per 100 g fresh granule content of the different brands was compared between the Zeus and (previously published data for) the Aisys. In vitro canister life of prefilled CO2 absorber canisters differed between brands, and depended on the amount of CO2 absorbent and chemical composition. Canister life expressed as FCU0.5 (the fraction of the canister used per hour) was proportional to FGF over 0.2-2 L min-1 range only, but was non-linear with higher FGF: FCU0.5 was larger than expected with FGF > 2 L min-1 , and even with FGF > minute ventilation FCU0.5 did not become zero, indicating some CO2 was being absorbed. Canister life on a per weight basis of the same brand is higher with the Zeus than the Aisys. Canister life of prefilled CO2 absorber canisters differs between brands. The FCU0.5 -FGF relationship is not linear across the entire FGF range. Canister life of prepacks of the same brand for the Zeus and Aisys differs, the exact etiology of which is probably multifactorial, and may include differences in the absolute amount of absorbent and different rebreathing characteristics of the machines.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app