Measured and Monte Carlo simulated surface dose reduction for superficial X-rays incident on tissue with underlying air or bone.

Medical Physics 2018 Februrary
PURPOSE: Measurement of surface dose reduction effects for superficial x-rays incident on tissue with underlying air or bone and comparison with Monte Carlo simulations of such effects. Further to investigate the correlation between surface dose reduction and changes in Compton backscatter spectra with tissue-bone separation.

METHODS: An Advanced Markus chamber with entrance window facing downstream on the surface of a solid water phantom was used to investigate changes in surface dose with an underlying air or bone interface located at various depths below the surface. Chamber readings were obtained for interface depths ranging from 1 to 100 mm using the 50 kV, 100 kV and 150 kV beams of an Xstrahl 150 x-ray unit, with field diameters (ϕ) = 2.5 cm and 5 cm. For each beam quality and field size the dose correction factor, DCF(t), namely the ratio of measured dose (t) to dose (t = 100 mm) was determined. Monte Carlo simulations of DCF(t) for air and bone interfaces in tissue are used to validate corresponding measured data. For a given beam and field size, the difference between simulated spectra with an air or bone interface at t = 3 mm was used to determine the Compton backscatter from bone at the surface.

RESULTS: For air, DCF(t < 40 mm) is less than unity and for a given t DCF(t) decreases as beam energy increases from 50 kV to 150 kV. Conversely for bone DCF(t < 40 mm) increases for a given t with increasing beam energy. In particular for t = 1 mm, ϕ = 5 cm, DCF for air(bone) are 0.90(0.92) and 0.86(0.96) for 50 kV and 150 kV, respectively. For the same tube potentials corresponding factors, ϕ = 2.5 cm, for air(bone) are 0.94(0.96) and 0.92(0.99). Calculated DCF(t) based on Monte Carlo simulations are consistent with experimental observations to within 2%. Monte Carlo simulations of x-ray spectra demonstrate the presence of Compton backscatter from underlying bone in tissue. With bone at 3 mm depth calculated backscatter spectra at the tissue surface suggest that surface dose is influenced by the proximity of bone and that this effect depends on beam quality.

CONCLUSIONS: This work demonstrates the feasibility of using an Advanced Markus chamber with entrance window facing downstream to investigate surface dose reduction with underlying air or bone in tissue. As the field size decreases and beam quality increases surface dose with underlying bone tends to full backscatter values even though tissue thicknesses are below those normally associated with full backscatter. Conversely with underlying bone close to the surface dose will increasingly fall below full backscatter values as the beam energy is reduced and field size is increased.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app