Reproducibility of relaxometry of human lumbar vertebrae at 3 Tesla using 1 H MR spectroscopy.

BACKGROUND: MR spectroscopy is widely used for fat fraction quantification of human lumbar vertebrae. However, the measurements need to be corrected for relaxation effects.

PURPOSE: The aim of this study was to determine the reproducibility of relaxometry in human lumbar vertebrae required for the correction of fat fraction measurements using magnetic resonance spectroscopy at 3 Tesla. Such information provides error estimates and guidance regarding reliability for future studies.

STUDY TYPE: Prospective.

SUBJECTS: Forty-six healthy volunteers (22 female [f], 24 male [m]) participated in this study.

FIELD STRENGTH: All subjects underwent three consecutive multi-TE /multi-TR MR spectroscopy measurements at 3 Tesla.

ASSESSMENT: A total of 2580 spectra of lumbar vertebrae L2 and L3 of 43 subjects (21f, 22m) were quantified using jMRUI software. Data were exported and mono-exponential fits were applied to the signals of water and fat compartments to derive relaxation times and calculate the fat fraction corrected for relaxation effects. Finally, relaxation times and fat fraction results of repeated measurements were analyzed for reproducibility.

STATISTICAL TESTS: Reproducibility was evaluated by calculating the coefficient of variation (CV). Influences of volunteer age and sex were tested by analysis of covariance.

RESULTS: The CV for all calculated parameters ranged between 1.22% (T2 of the fat compartment) and 3.02% (T1 of the fat compartment). Relaxation times and fat fraction were statistically different for female and male volunteers (P < 0.01) and relaxation times of the water compartment showed significant (P < 0.01) correlation with the fat fraction.

DATA CONCLUSION: Based on repeated acquisitions using the measurement parameters applied in this study, magnetic resonance spectroscopy allowed a reproducible calculation of the fat fraction corrected for relaxation effects. T1 of the water compartment showed high reproducibility and correlation with the fat fraction. It, therefore, might be considered as a parameter linked to the composition of the water compartment and patient health.

LEVEL OF EVIDENCE: 1 Technical Efficacy Stage 1 J. Magn. Reson. Imaging 2017.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app