Quantifying the Effect of 3T Magnetic Resonance Imaging Residual System Distortions and Patient-Induced Susceptibility Distortions on Radiation Therapy Treatment Planning for Prostate Cancer.
International Journal of Radiation Oncology, Biology, Physics 2018 Februrary 2
PURPOSE: To investigate the effect of magnetic resonance system- and patient-induced susceptibility distortions from a 3T scanner on dose distributions for prostate cancers.
METHODS AND MATERIALS: Combined displacement fields from the residual system and patient-induced susceptibility distortions were used to distort 17 prostate patient CT images. VMAT dose plans were initially optimized on distorted CT images and the plan parameters transferred to the original patient CT images to calculate a new dose distribution.
RESULTS: Maximum residual mean distortions of 3.19 mm at a radial distance of 25 cm and maximum mean patient-induced susceptibility shifts of 5.8 mm were found using the lowest bandwidth of 122 Hz per pixel. There was a dose difference of <0.5% between distorted and undistorted treatment plans. The 90% confidence intervals of the mean difference between the dCT and CT treatment plans were all within an equivalence interval of (-0.5, 0.5) for all investigated plan quality measures.
CONCLUSIONS: Patient-induced susceptibility distortions at high field strengths in closed bore magnetic resonance scanners are larger than residual system distortions after using vendor-supplied 3-dimensional correction for the delineated regions studied. However, errors in dose due to disturbed patient outline and shifts caused by patient-induced susceptibility effects are below 0.5%.
METHODS AND MATERIALS: Combined displacement fields from the residual system and patient-induced susceptibility distortions were used to distort 17 prostate patient CT images. VMAT dose plans were initially optimized on distorted CT images and the plan parameters transferred to the original patient CT images to calculate a new dose distribution.
RESULTS: Maximum residual mean distortions of 3.19 mm at a radial distance of 25 cm and maximum mean patient-induced susceptibility shifts of 5.8 mm were found using the lowest bandwidth of 122 Hz per pixel. There was a dose difference of <0.5% between distorted and undistorted treatment plans. The 90% confidence intervals of the mean difference between the dCT and CT treatment plans were all within an equivalence interval of (-0.5, 0.5) for all investigated plan quality measures.
CONCLUSIONS: Patient-induced susceptibility distortions at high field strengths in closed bore magnetic resonance scanners are larger than residual system distortions after using vendor-supplied 3-dimensional correction for the delineated regions studied. However, errors in dose due to disturbed patient outline and shifts caused by patient-induced susceptibility effects are below 0.5%.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app