We have located links that may give you full text access.
Journal Article
Multicenter Study
Randomized Controlled Trial
Effects of Rate on Analgesia in Kilohertz Frequency Spinal Cord Stimulation: Results of the PROCO Randomized Controlled Trial.
OBJECTIVE: The PROCO RCT is a multicenter, double-blind, crossover, randomized controlled trial (RCT) that investigated the effects of rate on analgesia in kilohertz frequency (1-10 kHz) spinal cord stimulation (SCS).
MATERIALS AND METHODS: Patients were implanted with SCS systems and underwent an eight-week search to identify the best location ("sweet spot") of stimulation at 10 kHz within the searched region (T8-T11). An electronic diary (e-diary) prompted patients for pain scores three times per day. Patients who responded to 10 kHz per e-diary numeric rating scale (ED-NRS) pain scores proceeded to double-blind rate randomization. Patients received 1, 4, 7, and 10 kHz SCS at the same sweet spot found for 10 kHz in randomized order (four weeks at each frequency). For each frequency, pulse width and amplitude were titrated to optimize therapy.
RESULTS: All frequencies provided equivalent pain relief as measured by ED-NRS (p ≤ 0.002). However, mean charge per second differed across frequencies, with 1 kHz SCS requiring 60-70% less charge than higher frequencies (p ≤ 0.0002).
CONCLUSIONS: The PROCO RCT provides Level I evidence for equivalent pain relief from 1 to 10 kHz with appropriate titration of pulse width and amplitude. 1 kHz required significantly less charge than higher frequencies.
MATERIALS AND METHODS: Patients were implanted with SCS systems and underwent an eight-week search to identify the best location ("sweet spot") of stimulation at 10 kHz within the searched region (T8-T11). An electronic diary (e-diary) prompted patients for pain scores three times per day. Patients who responded to 10 kHz per e-diary numeric rating scale (ED-NRS) pain scores proceeded to double-blind rate randomization. Patients received 1, 4, 7, and 10 kHz SCS at the same sweet spot found for 10 kHz in randomized order (four weeks at each frequency). For each frequency, pulse width and amplitude were titrated to optimize therapy.
RESULTS: All frequencies provided equivalent pain relief as measured by ED-NRS (p ≤ 0.002). However, mean charge per second differed across frequencies, with 1 kHz SCS requiring 60-70% less charge than higher frequencies (p ≤ 0.0002).
CONCLUSIONS: The PROCO RCT provides Level I evidence for equivalent pain relief from 1 to 10 kHz with appropriate titration of pulse width and amplitude. 1 kHz required significantly less charge than higher frequencies.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app