B-mode ultrasound assessment of pupillary function: Feasibility, reliability and normal values.
PloS One 2017
PURPOSE: To evaluate B-mode ultrasound as a novel method for the examination of pupillary function and to provide normal values for the pupillary reflex as assessed by B-mode ultrasound.
METHODS: 100 subjects (49 female, 51 male, mean [range] age 51 [18-80 years]) with no history of ophthalmologic disease, no clinically detectable pupillary defects, and corrected visual acuity ≥ 0.8 were included in this prospective observational study. B-mode ultrasound was performed with the subjects eyes closed using an Esaote-Mylab25 system according to current guidelines for orbital insonation. A standardized light stimulus was applied.
RESULTS: The mean ± standard deviation left and right pupillary diameters (PD) at rest were 4.7 ± 0.8 and 4.5 ± 0.8 mm. Following an ipsilateral light stimulus (Lstim), left and right constricted PD were 2.8 ± 0.6 and 2.7 ± 0.6 mm. Following a contralateral Lstim, left and right constricted PD were 2.7 ± 0.6 and 2.6 ± 0.5 mm. Left and right pupillary constriction time (PCT) following ipsilateral Lstim were 970 ± 261.6 and 967 ± 220 ms. Left and right PCT following a contralateral Lstim were 993.8 ± 192.6 and 963 ± 189.4 ms. Patient age was inversely correlated with PD at rest and with PD after ipsilateral and contralateral Lstim (all p<0.001), but not with PCT.
CONCLUSIONS: B-mode ultrasound is a simple, rapid and objective method for the quantitative assessment of pupillary function, which may prove useful in a variety of settings where eyelid retraction is impeded or an infrared pupillometry device is unavailable.
METHODS: 100 subjects (49 female, 51 male, mean [range] age 51 [18-80 years]) with no history of ophthalmologic disease, no clinically detectable pupillary defects, and corrected visual acuity ≥ 0.8 were included in this prospective observational study. B-mode ultrasound was performed with the subjects eyes closed using an Esaote-Mylab25 system according to current guidelines for orbital insonation. A standardized light stimulus was applied.
RESULTS: The mean ± standard deviation left and right pupillary diameters (PD) at rest were 4.7 ± 0.8 and 4.5 ± 0.8 mm. Following an ipsilateral light stimulus (Lstim), left and right constricted PD were 2.8 ± 0.6 and 2.7 ± 0.6 mm. Following a contralateral Lstim, left and right constricted PD were 2.7 ± 0.6 and 2.6 ± 0.5 mm. Left and right pupillary constriction time (PCT) following ipsilateral Lstim were 970 ± 261.6 and 967 ± 220 ms. Left and right PCT following a contralateral Lstim were 993.8 ± 192.6 and 963 ± 189.4 ms. Patient age was inversely correlated with PD at rest and with PD after ipsilateral and contralateral Lstim (all p<0.001), but not with PCT.
CONCLUSIONS: B-mode ultrasound is a simple, rapid and objective method for the quantitative assessment of pupillary function, which may prove useful in a variety of settings where eyelid retraction is impeded or an infrared pupillometry device is unavailable.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app