Multicenter Evaluation of a New Electrochemiluminescence Immunoassay for Everolimus Concentrations in Whole Blood.

BACKGROUND: The precise monitoring of everolimus, an immunosuppressant drug, is vital for transplant recipients due to its narrow therapeutic range. This study evaluated the analytical performance of a new electrochemiluminescence immunoassay (ECLIA) for everolimus concentrations in whole blood.

METHODS: Accuracy, imprecision, and sensitivity studies for the Roche Elecsys everolimus ECLIA were performed at 5 European laboratories. The ECLIA was compared with liquid chromatography-tandem mass spectrometry (LC-MS/MS) methods, as well as the Quantitative Microsphere System everolimus assay.

RESULTS: Everolimus ECLIA accuracies were within the range 100% ± 9%. Coefficients of variation (CVs) across the target range were ≤4.8% for repeatability and ≤8.4% for intermediate imprecision, whereas multisite reproducibility at lower (2.71 mcg/L) and higher everolimus concentrations (3.0-30.0 mcg/L) resulted in CVs of ≤13.7% and ≤12.4%, respectively. The CV at the assay's lower limit of quantification without considering bias was excellent, estimated as ≤9.3% at 0.5 mcg/L. The weighted Deming regression analysis, used for comparison of the results obtained by everolimus ECLIA and by LC-MS/MS methods, yielded a slope of 1.21 [95% confidence interval (CI): 1.15-1.26], intercept of 0.478 mcg/L (95% CI: 0.241-0.716), and a Pearson correlation coefficient (r) of 0.91. A single-site comparison between the ECLIA and the Quantitative Microsphere System assay revealed a slope of 1.05 (95% CI: 0.917-1.17), intercept of 1.03 mcg/L (95% CI: 0.351-1.70), and r of 0.91.

CONCLUSIONS: Based on these results, the Roche Elecsys everolimus ECLIA can be considered suitable for routine therapeutic drug monitoring. A positive bias was observed with respect to LC-MS/MS methods, suggesting that it may be necessary to rebaseline individual patients when switching from LC-MS/MS to the ECLIA; however, this must also be considered for any change of method for everolimus measurement.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app