We have located links that may give you full text access.
JOURNAL ARTICLE
VALIDATION STUDY
Comprehensive clinical commissioning and validation of the RayStation treatment planning system for proton therapy with active scanning and passive treatment techniques.
Physica Medica : PM 2017 November
PURPOSE: To commission the treatment planning system (TPS) RayStation for proton therapy including beam models for spot scanning and for uniform scanning.
METHODS: Tests consist of procedures from ESTRO booklet number 7, the German DIN for constancy checks of TPSs, and extra tests checking the dose perturbation function. The dose distributions within patients were verified in silico by a comparison of 65 clinical treatment plans with the TPS XiO. Dose-volume parameters, dose differences, and three-dimensional gamma-indices serve as measures of similarity. The monthly constancy checks of Raystation have been automatized with a script.
RESULTS: The basic functionality of the software complies with ESTRO booklet number 7. For a few features minor enhancements are suggested. The dose distribution in RayStation agrees with the calculation in XiO. This is supported by a gamma-index (3mm/3%) pass rate of >98.9% (median over 59 plans) for the volume within the 20% isodose line and a difference of <0.3% of V95 of the PTV (median over 59 plans). If spot scanning is used together with a range shifter, the dose level calculated by RayStation can be off by a few percent.
CONCLUSIONS: RayStation can be used for the creation of clinical proton treatment plans. Compared to XiO RayStation has an improved modelling of the lateral dose fall-off in passively delivered fields. For spot scanning fields with range shifter blocks an empirical adjustment of monitor units is required. The computation of perturbed doses also allows the evaluation of the robustness of a treatment plan.
METHODS: Tests consist of procedures from ESTRO booklet number 7, the German DIN for constancy checks of TPSs, and extra tests checking the dose perturbation function. The dose distributions within patients were verified in silico by a comparison of 65 clinical treatment plans with the TPS XiO. Dose-volume parameters, dose differences, and three-dimensional gamma-indices serve as measures of similarity. The monthly constancy checks of Raystation have been automatized with a script.
RESULTS: The basic functionality of the software complies with ESTRO booklet number 7. For a few features minor enhancements are suggested. The dose distribution in RayStation agrees with the calculation in XiO. This is supported by a gamma-index (3mm/3%) pass rate of >98.9% (median over 59 plans) for the volume within the 20% isodose line and a difference of <0.3% of V95 of the PTV (median over 59 plans). If spot scanning is used together with a range shifter, the dose level calculated by RayStation can be off by a few percent.
CONCLUSIONS: RayStation can be used for the creation of clinical proton treatment plans. Compared to XiO RayStation has an improved modelling of the lateral dose fall-off in passively delivered fields. For spot scanning fields with range shifter blocks an empirical adjustment of monitor units is required. The computation of perturbed doses also allows the evaluation of the robustness of a treatment plan.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app