A Protective Eye Shield Reduces Limbal Strain and Its Variability During Simulated Sleep in Adults With Glaucoma.
Journal of Glaucoma 2018 January
PURPOSE: To determine the effect of wearing a protective eye shield (mask) on limbal strain magnitude and variability in glaucoma eyes when sleeping with 1 side of the face down (FD) against a pillow.
METHODS: A prospective, randomized, interventional trial was conducted at the Wilmer Eye Institute with 36 glaucoma patients. A contact lens sensor measured limbal strain (output in equivalent millivolts) during intervals of up to 60 minutes in lateral decubitus, FD, and supine positions. Eighteen subjects wore a mask during 1 of 2 FD intervals, with randomized assignment of the interval. Data from additional trials with no mask were included in some analyses. In addition, some facial-feature dimensions from 3D scanned images of 23 subjects were compared with limbal strain data.
RESULTS: Wearing a mask trends toward a reduced mean change in contact lens sensor output (limbal strain) on moving to a FD positions [+34.1 mVeq, P=0.01 reduced by -22.3 mVeq, P=0.09 (n=36)]. Mask wearing reduced variability in strain while FD [-22.8 mVeq, P=0.04 (n=18)]. In eyes with past progressive visual field loss, the effect of the mask reduced mean strain change when moving to FD [-44.8 mVeq, P=0.02 (n=31)]. Longer corneal apex to nose-tip and to temple lengths were associated with reduced variability while FD [P=0.02 and 0.04, respectively (n=23)]. Treating both lengths as confounding factors increased statistical significance, particularly for analysis of the no-mask change in strain data moving to and from the FD position [P=0.004 to 0.002 and P=0.03 to 0.01 (n=23)].
CONCLUSION AND RELEVANCE: Wearing a mask reduced limbal strain and variation in limbal strain during simulated FD sleep, particularly in eyes with past field worsening, as did some facial features.
METHODS: A prospective, randomized, interventional trial was conducted at the Wilmer Eye Institute with 36 glaucoma patients. A contact lens sensor measured limbal strain (output in equivalent millivolts) during intervals of up to 60 minutes in lateral decubitus, FD, and supine positions. Eighteen subjects wore a mask during 1 of 2 FD intervals, with randomized assignment of the interval. Data from additional trials with no mask were included in some analyses. In addition, some facial-feature dimensions from 3D scanned images of 23 subjects were compared with limbal strain data.
RESULTS: Wearing a mask trends toward a reduced mean change in contact lens sensor output (limbal strain) on moving to a FD positions [+34.1 mVeq, P=0.01 reduced by -22.3 mVeq, P=0.09 (n=36)]. Mask wearing reduced variability in strain while FD [-22.8 mVeq, P=0.04 (n=18)]. In eyes with past progressive visual field loss, the effect of the mask reduced mean strain change when moving to FD [-44.8 mVeq, P=0.02 (n=31)]. Longer corneal apex to nose-tip and to temple lengths were associated with reduced variability while FD [P=0.02 and 0.04, respectively (n=23)]. Treating both lengths as confounding factors increased statistical significance, particularly for analysis of the no-mask change in strain data moving to and from the FD position [P=0.004 to 0.002 and P=0.03 to 0.01 (n=23)].
CONCLUSION AND RELEVANCE: Wearing a mask reduced limbal strain and variation in limbal strain during simulated FD sleep, particularly in eyes with past field worsening, as did some facial features.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app