In vitro trials of a wearable artificial kidney (WAK).

BACKGROUND: This paper presents results of in vitro trials of a wearable artificial kidney (WAK).

MATERIALS AND METHODS: To carry out the trials, a test bench was developed. It consisted of a patient imitator, a tubing set and measuring equipment.

RESULTS: Our WAK can eliminate urea, uric acid and creatinine from spent dialysate with mean mass rates of 0.85 ± 0.1 g/h, 0.10 ± 0.04 g/h and 0.05 ± 0.01 g/h, respectively. Concentrations of Na+, Cl- and Ca2+ ions were kept in the 10% range from initial values. We demonstrated that our WAK can operate for 38 hours without having to replace expendable materials.

CONCLUSIONS: The WAK that we developed can eliminate metabolites from spent dialysis fluid with mass rates that are sufficient to maintain stable, physiologically normal metabolite concentrations in patients' blood.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app