Evaluation of AAPM Reports 204 and 220: Estimation of effective diameter, water-equivalent diameter, and ellipticity ratios for chest, abdomen, pelvis, and head CT scans.
Journal of Applied Clinical Medical Physics 2018 January
PURPOSE: To confirm AAPM Reports 204/220 and provide data for the future expansion of these reports by: (a) presenting the first large-scale confirmation of the reports using clinical data, (b) providing the community with size surrogate data for the head region which was not provided in the original reports, and additionally providing the measurements of patient ellipticity ratio for different body regions.
METHOD: A total of 884 routine scans were included in our analysis including data from the head, thorax, abdomen, and pelvis for adults and pediatrics. We calculated the ellipticity ratio and all of the size surrogates presented in AAPM Reports 204/220. We correlated the purely geometric-based metrics with the "gold standard" water-equivalent diameter (DW ).
RESULTS: Our results and AAPM Reports 204/220 agree within our data's 95% confidence intervals. Outliers to the AAPM reports' methods were caused by excess gas in the GI tract, exceptionally low BMI, and cranial metaphyseal dysplasia. For the head, we show lower correlation (R2 = 0.812) between effective diameter and DW relative to other body regions. The ellipticity ratio of the shoulder region was the highest at 2.28 ± 0.22 and the head the smallest at 0.85 ± 0.08. The abdomen pelvis, chest, thorax, and abdomen regions all had ellipticity values near 1.5.
CONCLUSION: We confirmed AAPM reports 204/220 using clinical data and identified patient conditions causing discrepancies. We presented new size surrogate data for the head region and for the first time presented ellipticity data for all regions. Future automatic exposure control characterization should include ellipticity information.
METHOD: A total of 884 routine scans were included in our analysis including data from the head, thorax, abdomen, and pelvis for adults and pediatrics. We calculated the ellipticity ratio and all of the size surrogates presented in AAPM Reports 204/220. We correlated the purely geometric-based metrics with the "gold standard" water-equivalent diameter (DW ).
RESULTS: Our results and AAPM Reports 204/220 agree within our data's 95% confidence intervals. Outliers to the AAPM reports' methods were caused by excess gas in the GI tract, exceptionally low BMI, and cranial metaphyseal dysplasia. For the head, we show lower correlation (R2 = 0.812) between effective diameter and DW relative to other body regions. The ellipticity ratio of the shoulder region was the highest at 2.28 ± 0.22 and the head the smallest at 0.85 ± 0.08. The abdomen pelvis, chest, thorax, and abdomen regions all had ellipticity values near 1.5.
CONCLUSION: We confirmed AAPM reports 204/220 using clinical data and identified patient conditions causing discrepancies. We presented new size surrogate data for the head region and for the first time presented ellipticity data for all regions. Future automatic exposure control characterization should include ellipticity information.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app