[Bruch's membrane opening minimum rim width : Correlation and diagnostic accuracy in comparison to peripapillary retinal nerve fiber layer thickness].

BACKGROUND: Glaucoma is one of the main causes of blindness in the Western hemisphere. Because the disease often painlessly progresses it remains unnoticed until major optic nerve head damage occurs in many cases. That is why new, more sensitive diagnostic methods are needed. Bruch's membrane opening minimum rim width (BMO-MRW), measured with the new glaucoma module premium edition (GMPE) was recently introduced as a more accurate tool to detect glaucomatous changes. The purpose of this study was to assess the correlation of SPECTRALIS® spectral domain optical coherence tomography (SD-OCT) and the GMPE anatomic positioning module (APS module) for retinal nerve fiber layer thickness (RNFLT) measurements. The second aim was to assess the diagnostic accuracy of BMO-MRW.

METHODS: Prospective study of 41 eyes (41 patients) with glaucoma and 26 eyes from 26 healthy controls. Scans were obtained using SPECTRALIS® SD-OCT and RNFLT was measured with both modules and compared using Spearman's rank test. The BMO-MRW was assessed by GMPE. Sensitivity, specificity and area under receiver operating characteristics curves (AUROC) of each sector of the optic nerve were calculated and compared using the method of Delong et al.

RESULTS: We found a positive correlation (0.694-0.955, p < 0.0001) between RNFLT measurements by standard SD-OCT and all diameters of RNFLT of APS module within all sectors. The AUROC of RNFLT in standard SD-OCT was 0.693 for the inferior nasal sector (NI) and BMO-MRW was 0.85 in NI. The difference in AUROC was statistically significant (p = 0.0049). No other sector showed statistically significant differences.

CONCLUSION: The RNFLT measurements of both modules showed a positive correlation and appear to be comparable. The BMO-MRW in one sector (NI) showed a significantly higher accuracy of measurement than standard RNFLT. All other sectors showed a comparable accuracy of measurement.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app