We have located links that may give you full text access.
COMPARATIVE STUDY
EVALUATION STUDY
JOURNAL ARTICLE
Real-time two-dimensional Shear-wave elastography for liver stiffness in children: Interobserver variation and effect of breathing technique.
European Journal of Radiology 2017 December
OBJECTIVES: To evaluate interobserver variation and to assess the effect of the breathing technique on liver stiffness (LS) measurement using real-time two-dimensional shear-wave elastography (2D-SWE) in children.
METHODS: Patients who underwent real-time 2D-SWE by two operators (group 1) and patients who underwent real-time 2D-SWE by one operator using the free-breathing and breath-holding techniques (group 2) were included. In group 1, interobserver agreement was assessed using the intraclass correlation coefficient (ICC) and the 95% limits-of-agreement was analyzed to evaluate the maximum change in the LS values based on Bland-Altman analysis. In group 2, the Bland-Altman plot and the paired t-test were used to determine the effect of breathing technique on LS measurement.
RESULTS: The interobserver agreement of the LS measurement in group 1 (n=63) was excellent (ICC, 0.953), and the 95% limits-of-agreement was 31.8% of the mean LS values. In group 2 (n=45), the free-breathing technique revealed systematically lower LS with a mean difference of -11.1% of the mean LS values, compared to the breath-holding technique (P=0.016).
CONCLUSION: LS measurement assessed by real-time 2D-SWE in children showed excellent interobserver agreement. Changes in LS values less than 31.8% of the mean LS values may represent the measurement error. The free-breathing technique showed lower LS values, compared to the breath-holding technique.
METHODS: Patients who underwent real-time 2D-SWE by two operators (group 1) and patients who underwent real-time 2D-SWE by one operator using the free-breathing and breath-holding techniques (group 2) were included. In group 1, interobserver agreement was assessed using the intraclass correlation coefficient (ICC) and the 95% limits-of-agreement was analyzed to evaluate the maximum change in the LS values based on Bland-Altman analysis. In group 2, the Bland-Altman plot and the paired t-test were used to determine the effect of breathing technique on LS measurement.
RESULTS: The interobserver agreement of the LS measurement in group 1 (n=63) was excellent (ICC, 0.953), and the 95% limits-of-agreement was 31.8% of the mean LS values. In group 2 (n=45), the free-breathing technique revealed systematically lower LS with a mean difference of -11.1% of the mean LS values, compared to the breath-holding technique (P=0.016).
CONCLUSION: LS measurement assessed by real-time 2D-SWE in children showed excellent interobserver agreement. Changes in LS values less than 31.8% of the mean LS values may represent the measurement error. The free-breathing technique showed lower LS values, compared to the breath-holding technique.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app