Add like
Add dislike
Add to saved papers

Noise reduction angiographic imaging technology reduces radiation dose during bronchial artery embolization.

PURPOSE: Comparison of radiation doses in patients undergoing angiographic bronchial artery embolization (BAE) before and after a noise reduction imaging technology upgrade.

METHODS: We performed a retrospective study of 70 patients undergoing BAE. Procedures were performed before (n=32) and after (n=38) the technology upgrade containing additional filters and improved image-processing. Cumulative air kerma (AK), cumulative dose area product (DAP), number of exposure frames, total fluoroscopy time and amount of contrast agent were recorded. Mean values were calculated and compared using two-tailed t-tests. DSA image quality was assessed independently by two blinded readers and compared using the Wilcoxon signed-rank test.

RESULTS: Using the new technology resulted in a significant reduction of 59% in DAP (149.2 (103.1-279.1) vs. 54.8 (38.2-100.7) Gy*cm2 , p<0.001) and a significant reduction of 60% for AK (1.3 (0.6-1.9) vs. 0.5 (0.3-0.9) Gy, p<0.001) in comparison to procedures before the upgrade. There was no significant difference between the number of exposure frames in both groups (251±181 vs. 254±133 frames, p=0.07), time of fluoroscopy (28.8 (18.5-50.4) vs. 28.1 (23.3-38.7) min, p=0.73), or the amount of contrast agent used (139.5±70.8 vs. 163.1±63.1ml, p=0.11). No significant difference regarding image quality could be detected (3 (2,3) vs. 3 (2-4), p=0.64).

CONCLUSIONS: The new angiographic noise reduction technology significantly decreases the radiation dose during bronchial artery embolization without compromising image quality or increasing time of fluoroscopy or contrast volume.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app