An evaluation of the SENTiFIT 270 analyser for quantitation of faecal haemoglobin in the investigation of patients with suspected colorectal cancer.
Clinical Chemistry and Laboratory Medicine : CCLM 2018 March 29
BACKGROUND: An evaluation of SENTiFIT® 270 (Sentinel Diagnostics, Italy; Sysmex, Spain) analyser for the quantitation of faecal haemoglobin (f-Hb) was performed.
METHODS: The analytical imprecision, linearity, carry over and f-Hb stability were determined. Evaluation of the diagnostic accuracy was performed on 487 patients.
RESULTS: Within-run and between-run imprecision ranged 1.7%-5.1% and 3.8%-6.2%, respectively. Linearity studies revealed a mean recovery of 101.1% (standard deviation, 6.7%) for all dilutions. No carry over was detected below 7650 μg Hb/g faeces. Decay of f-Hb in refrigerated samples ranged 0.2%-0.5% per day. f-Hb in patients with advanced colorectal neoplasia (ACRN) (colorectal cancer [CRC] plus advanced adenoma [AA]) were significantly higher than from those with a normal colonoscopy. Sensitivity for ACRN at f-Hb cutoffs from 10 to 60 μg Hb/g faeces ranged from 28.9% (95% confidence interval [CI], 21.7%-37.2%) to 46.5% (95% CI, 38.1%-55%), the specificity ranged from 85% (95% CI, 82.3%-87.3%) to 93.2% (95% CI, 91.2%-94.8%), positive predictive values for detecting CRC and AA ranged from 11.6% (95% CI, 7.6%-17.2%) to 20.6% (95% CI, 13.3%-30.3%) and from 34.7% (95% CI, 28.1%-42%) to 42.3% (95% CI, 32.4%-52.7%), respectively, and the negative predictive value for ACRN ranged from 90.2% (95% CI, 87.9%-92.2%) to 88.4% (95% CI, 86%-90.4%). Using two samples per patient sensitivity increased with a slight decrease in specificity.
CONCLUSIONS: The analytical and clinical performances of SENTiFIT assay demonstrate a specific and accurate test for detecting ACRN in symptomatic patients and those undergoing surveillance.
METHODS: The analytical imprecision, linearity, carry over and f-Hb stability were determined. Evaluation of the diagnostic accuracy was performed on 487 patients.
RESULTS: Within-run and between-run imprecision ranged 1.7%-5.1% and 3.8%-6.2%, respectively. Linearity studies revealed a mean recovery of 101.1% (standard deviation, 6.7%) for all dilutions. No carry over was detected below 7650 μg Hb/g faeces. Decay of f-Hb in refrigerated samples ranged 0.2%-0.5% per day. f-Hb in patients with advanced colorectal neoplasia (ACRN) (colorectal cancer [CRC] plus advanced adenoma [AA]) were significantly higher than from those with a normal colonoscopy. Sensitivity for ACRN at f-Hb cutoffs from 10 to 60 μg Hb/g faeces ranged from 28.9% (95% confidence interval [CI], 21.7%-37.2%) to 46.5% (95% CI, 38.1%-55%), the specificity ranged from 85% (95% CI, 82.3%-87.3%) to 93.2% (95% CI, 91.2%-94.8%), positive predictive values for detecting CRC and AA ranged from 11.6% (95% CI, 7.6%-17.2%) to 20.6% (95% CI, 13.3%-30.3%) and from 34.7% (95% CI, 28.1%-42%) to 42.3% (95% CI, 32.4%-52.7%), respectively, and the negative predictive value for ACRN ranged from 90.2% (95% CI, 87.9%-92.2%) to 88.4% (95% CI, 86%-90.4%). Using two samples per patient sensitivity increased with a slight decrease in specificity.
CONCLUSIONS: The analytical and clinical performances of SENTiFIT assay demonstrate a specific and accurate test for detecting ACRN in symptomatic patients and those undergoing surveillance.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app