Field-size correction factors of a radiophotoluminescent glass dosimeter for small-field and intensity-modulated radiation therapy beams.

Medical Physics 2018 January
PURPOSE: We evaluated the energy responses of a radiophotoluminescent glass dosimeter (RPLD) to variations in small-field and intensity-modulated radiation therapy (IMRT) conditions using experimental measurements and Monte Carlo simulation.

METHODS: Several sizes of the jaw and multileaf collimator fields and various plan-class IMRT-beam measurements were performed using the RPLD and an ionization chamber. The field-size correction factor for the RPLD was determined for 6- and 10-MV x rays. This correction factor, together with the perturbation factor, was also calculated using Monte Carlo simulation with the EGSnrc/egs_chamber user code. In addition, to evaluate the response of the RPLD to clinical-class-specific reference fields, the field-size correction factor for the clinical IMRT plan was measured.

RESULTS: The calculated field-size correction factor ranged from 1.007 to 0.981 (for 6-MV x rays) and from 1.012 to 0.990 (for 10-MV x rays) as the jaw-field size ranged from 1 × 1 cm2 to 20 × 20 cm2 . The atomic composition perturbation factor for these jaw fields decreased by 3.2% and 1.9% for the 6- and 10-MV fields, respectively. The density perturbation factor was unity for field sizes ranging from 3 × 3 cm2 to 20 × 20 cm2 , whereas that for field sizes ranging from 3 × 3 cm2 to 1 × 1 cm2 decreased by 3.2% (for 6-MV x rays) and 4.3% (for 10-MV x rays). The volume-averaging factor rapidly increased for field sizes below 1.6 × 1.6 cm2 . The results for the MLC fields were similar to those for the jaw fields. For plan-class IMRT beams, the field-size correction and perturbation factors were almost unity. The difference between the doses measured using the RPLD and ionization chamber was within 1.2% for the clinical IMRT plan at the planning-target volume (PTV) region.

CONCLUSIONS: For small fields of size 1.6 × 1.6 cm2 or less, it was clarified that the volume averaging and density perturbation were the dominant effects responsible for the variation in the RPLD response. Moreover, perturbation correction is required when measuring a field size 1.0 × 1.0 cm2 or less. Under the IMRT conditions, the difference in the responses of the RPLD between the reference conditions and the PTV region calculated by Monte Carlo simulation did not exceed 0.8%. These results indicate that it is feasible to measure IMRT dosage using an RPLD at the PTV region.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app