Bore Disruption: An Unusual Mechanical Failure of Two Hip Hemiarthroplasties.

Joints 2017 March
We report a case of two hemiarthroplasties (HAs) that underwent an unexpected mechanical failure. Two patients affected by femoral fractures were treated with an HA. At 5 and 7 years after surgery, they showed a breakage of the HA at the head-stem junction. We analyzed macroscopically and microscopically (by a scanning electron microscopy) the surfaces of the broken prostheses. In both cases, a fracture was detected starting at the joining point between the femoral head and the slot that receives the taper cone of the stem (the so-called bore), with the typical macroscopic and microscopic pattern of a fatigue fracture. The fatigue fracture resulted from repeated cycles of torsional stresses throughout the years. In literature, we did not find any other report of such HA failure.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app