Improved clinical workflow for simultaneous whole-body PET/MRI using high-resolution CAIPIRINHA-accelerated MR-based attenuation correction.

PURPOSE: To explore the value and reproducibility of a novel magnetic resonance based attenuation correction (MRAC) using a CAIPIRINHA-accelerated T1-weighted Dixon 3D-VIBE sequence for whole-body PET/MRI compared to the clinical standard.

METHODS: The PET raw data of 19 patients from clinical routine were reconstructed with standard MRAC (MRACstd ) and the novel MRAC (MRACcaipi ), a prototype CAIPIRINHA accelerated Dixon 3D-VIBE sequence, both acquired in 19 s/bed position. Volume of interests (VOIs) for liver, lung and all voxels of the total image stack were created to calculate standardized uptake values (SUVmean ) followed by inter-method agreement (Passing-Bablok regression, Bland-Altman analysis). A voxel-wise SUV comparison per patient was performed for intra-individual correlation between MRACstd and MRACcaipi . Difference images (MRACstd -MRACcaipi ) of attenuation maps and SUV images were calculated. The image quality of in/opposed-phase water and fat images obtained from MRACcaipi was assessed by two readers on a 5-point Likert-scale including intra-class coefficients for inter-reader agreement.

RESULTS: SUVmean correlations of VOIs demonstrated high linearity (0.95<Spearman's rho<1, p<0.0001, respectively), substantiated by voxel-wise SUV scatter-plots (1.79×108 pixels). Outliers could be explained by different physiological conditions between the scans such as different segmentation of air-containing tissue, lungs, kidneys, metal implants, diaphragm edge or small air bubbles in the gastrointestinal tracts that moved between MRAC acquisitions. Nasal sinuses and the trachea were better segmented in MRACcaipi . High-resolution T1w Dixon 3D VIBE images were acquired in all cases and could be used for PET/MRI fusion. MRACcaipi images were of high diagnostic quality (4.2±0.8) with 0.92-0.96 intra-class correlation.

CONCLUSIONS: The novel prototype MRACcaipi extends the value for attenuation correction by providing a high spatial resolution DIXON-based dataset suited for diagnostic assessment towards time-efficient whole-body PET/MRI.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app