English Abstract
Journal Article
Add like
Add dislike
Add to saved papers

[Repeatability and variability of total T4 measurements at three German veterinary laboratories].

OBJECTIVE: To evaluate the reproducibility of serum testing for total thyroxine (T4) in three German laboratories.

MATERIALS AND METHODS: Serum was taken from 53 dogs with suspected hypothyroidism and divided into five aliquots. Three aliquots of each sample were marked with different names and sent to one laboratory (two aliquots simultaneously and one aliquot 1-3 days later). The remaining aliquots were sent to two other laboratories. Laboratory 1 used an enzyme immunoassay for T4 measurements, whereas laboratories 2 and 3 used a chemiluminescence immunoassay. The agreement between the three laboratories (values within or below the reference interval) was determined using the Cohen's Kappa test. The intra- and interassay variability was calculated for each laboratory and the agreement between samples submitted to the same laboratory was also determined using the Cohen's Kappa test.

RESULTS: For n = 23/41 patients tested simultaneously in the three laboratories, all three values were either uniformly below, within or above the respective reference interval. The Cohen's Kappa value for intra- and interassay agreement was 1.0 in laboratory 2 (n = 15, complete agreement), 0.33 in laboratory 1 (n = 16) and 0.37 (intra-) and 0.19 (interassay agreement) in laboratory 3 (n = 16, low agreement). There was a low agreement between laboratories 1 and 2 and between laboratories 1 and 3 (κ = 0.30 and 0.25, respectively), while a high agreement was determined between laboratories 2 and 3 (κ = 0.68). The intrassay variability of laboratories 1, 2 and 3 was 13.6%, 5.0% and 10.4%, the interassay variability 17.2%, 5.1% and 17.4%, respectively.

CONCLUSION AND CLINICAL RELEVANCE: The differences in the measurement of thyroxine concentrations of the same serum sample in different laboratories and at different time points in the same laboratory underline the high relevance of interpreting laboratory results in context with the clinical signs of hypothyroidism as well as other laboratory values such as TSH concentration.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app