Biocompatibility and degradation properties of WE43 Mg alloys with and without heat treatment: In vivo evaluation and comparison in a cranial bone sheep model.

PURPOSE: Orthopedic and maxillofacial bone fractures are routinely treated by titanium internal fixation, which may be prone to exposure, infection or intolerance. Magnesium (Mg) and its alloys represent promising alternatives to produce biodegradable osteosynthesis devices, with biocompatibility and, specifically, hydrogen gas production during the degradation process, being the main drawback. Aim of this study is to test and compare biocompatibility, degradation rate and physiscochemical properties of two Mg-alloys to identify which one possesses the most suitable characteristics to be used as resorbable hardware in load-bearing fracture sites.

MATERIALS AND METHODS: As-cast (WE43) and T5 Mg-alloys were tested for biocompatibility, physical, mechanical and degradation properties. Microstructure was assessed by optical microscopy, scanning electron microscopy (SEM) and energy-dispersive spectroscopy (EDS); mechanical properties were tested utilizing quasi-static compression and failure analysis. Locoregional biocompatibility was tested by sub-periosteal implantation on the fronto-nasal region of large-animal model (sheep): regional immunoreaction and metal accumulation was analyzed by LA-ICP of tributary lymph-nodes, local reactions were analyzed through histological preparation including bone, implant and surrounding soft tissue.

RESULTS: Mechanically, T5 alloy showed improvement in strength compared to the as-cast. Lymph-node Mg accumulation depicted no differences between control (no implant) and study animals. Both alloys showed good biocompatibility and osteogenesis-promoting properties.

CONCLUSION: This study demonstrated excellent biocompatibility and osteogenesis-promoting capabilities of the tested alloys, providing a platform for further studies to test them in a maxillofacial fracture setting. T-5 alloy displayed more stability and decreased degradation rate than the as-cast.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app