Journal Article
Research Support, N.I.H., Extramural
Add like
Add dislike
Add to saved papers

Intracochlear Measurements of Interaural Time and Level Differences Conveyed by Bilateral Bone Conduction Systems.

Otology & Neurotology 2017 December
HYPOTHESIS: Intracochlear pressures (PIC) and stapes velocity (Vstap) elicited by bilaterally placed bone-anchored hearing devices (BAHD) will be systematically modulated by imposed interaural time (ITD) and level differences (ILD), demonstrating the potential for users of bilateral BAHD to access these binaural cues.

BACKGROUND: BAHD are traditionally implanted unilaterally under the assumption that transcranial cross-talk limits interaural differences. Recent studies have demonstrated improvements in binaural and spatial performance with bilateral BAHD; however, objective measures of binaural cues from bilateral BAHDs are lacking.

METHODS: Bone-conduction transducers were coupled to both mastoids of cadaveric specimens via implanted titanium abutments. PIC and Vstap were measured using intracochlear pressure probes and laser Doppler vibrometry, respectively, during stimulation with pure-tone stimuli of varied frequency (250-4000 Hz) under ipsilateral, contralateral, and bilateral ITD (-1 to 1 ms) and ILD (-20 to 20 dB) conditions.

RESULTS: Bilateral stimulation produced constructive and destructive interference patterns that varied dramatically with ITD and stimulus frequency. Variation of ITD led to large variation of PIC and Vstap, with opposing effects in ipsilateral and contralateral ears expected to lead to "ITD to ILD conversion." Variation of ILD produced more straightforward (monotonic) variations of PIC and Vstap, with ipsilateral-favoring ILD producing higher PIC and Vstap than contralateral-favoring.

CONCLUSION: Variation of ITDs and ILDs conveyed by BAHDs systematically modulated cochlear inputs. While transcranial cross-talk leads to complex interactions that depend on cue type and stimulus frequency, binaural disparities potentiate binaural benefit, providing a basis for improved sound localization and speech-in-noise perception.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app