A Randomized Control Trial of Cardiopulmonary Feedback Devices and Their Impact on Infant Chest Compression Quality: A Simulation Study.

Pediatric Emergency Care 2017 October 28
OBJECTIVES: In effort to improve chest compression quality among health care providers, numerous feedback devices have been developed. Few studies, however, have focused on the use of cardiopulmonary resuscitation feedback devices for infants and children. This study evaluated the quality of chest compressions with standard team-leader coaching, a metronome (MetroTimer by ONYX Apps), and visual feedback (SkillGuide Cardiopulmonary Feedback Device) during simulated infant cardiopulmonary resuscitation.

METHODS: Seventy voluntary health care providers who had recently completed Pediatric Advanced Life Support or Basic Life Support courses were randomized to perform simulated infant cardiopulmonary resuscitation into 1 of 3 groups: team-leader coaching alone (control), coaching plus metronome, or coaching plus SkillGuide for 2 minutes continuously. Rate, depth, and frequency of complete recoil during cardiopulmonary resuscitation were recorded by the Laerdal SimPad device for each participant. American Heart Association-approved compression techniques were randomized to either 2-finger or encircling thumbs.

RESULTS: The metronome was associated with more ideal compression rate than visual feedback or coaching alone (104/min vs 112/min and 113/min; P = 0.003, 0.019). Visual feedback was associated with more ideal depth than auditory (41 mm vs 38.9; P = 0.03). There were no significant differences in complete recoil between groups. Secondary outcomes of compression technique revealed a difference of 1 mm. Subgroup analysis of male versus female showed no difference in mean number of compressions (221.76 vs 219.79; P = 0.72), mean compression depth (40.47 vs 39.25; P = 0.09), or rate of complete release (70.27% vs 64.96%; P = 0.54).

CONCLUSIONS: In the adult literature, feedback devices often show an increase in quality of chest compressions. Although more studies are needed, this study did not demonstrate a clinically significant improvement in chest compressions with the addition of a metronome or visual feedback device, no clinically significant difference in Pediatric Advanced Life Support-approved compression technique, and no difference between compression quality between genders.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app