Validation of a new piezoelectric device for noninvasive measurement of central aortic systolic blood pressure.
Blood Pressure Monitoring 2018 Februrary
OBJECTIVE: The objective of this study was to compare the aortic piezoelectric device for noninvasive measurement of central aortic systolic blood pressure (cSBP) with the SphygmoCor.
PARTICIPANTS AND METHODS: A total of 85 participants from both sexes, aged 18-80 years, were stratified into three age groups (<30, 30-60, >60 years), with an equal number of healthy volunteers and hypertensive patients. We performed three cSBP measurements with each device, in an alternate manner, using the Bland-Altman method to determine the level of agreement. The standard of the Association for the Advancement of Medical Instrumentation for brachial blood pressure evaluation was used for the comparison.
RESULTS: The mean cSBPs were 109.3±12.05 and 109.0±12.2 mmHg with the SphygmoCor and the Aortic device, respectively, showing a strong correlation (r=0.98, P<0.001). A mean difference of 0.35±2.43 mmHg (95% confidence interval: 0.17-0.87, P=NS) was obtained with the Bland-Altman method. The 95% limits of agreement was -4.4 to +5.1 mmHg.
CONCLUSION: Complying with the Association for the Advancement of Medical Instrumentation criteria, cSBP measurements obtained with the Aortic and the SphygmoCor devices are equivalent.
PARTICIPANTS AND METHODS: A total of 85 participants from both sexes, aged 18-80 years, were stratified into three age groups (<30, 30-60, >60 years), with an equal number of healthy volunteers and hypertensive patients. We performed three cSBP measurements with each device, in an alternate manner, using the Bland-Altman method to determine the level of agreement. The standard of the Association for the Advancement of Medical Instrumentation for brachial blood pressure evaluation was used for the comparison.
RESULTS: The mean cSBPs were 109.3±12.05 and 109.0±12.2 mmHg with the SphygmoCor and the Aortic device, respectively, showing a strong correlation (r=0.98, P<0.001). A mean difference of 0.35±2.43 mmHg (95% confidence interval: 0.17-0.87, P=NS) was obtained with the Bland-Altman method. The 95% limits of agreement was -4.4 to +5.1 mmHg.
CONCLUSION: Complying with the Association for the Advancement of Medical Instrumentation criteria, cSBP measurements obtained with the Aortic and the SphygmoCor devices are equivalent.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app