Contrast enhancement in abdominal computed tomography: influence of photon energy of different scanners.

OBJECTIVE: Different CT scanners have different X-ray spectra and photon energies indicating that contrast enhancement vary among scanners. However, this issue has not been fully validated; therefore, we performed phantom and clinical studies to assess this difference.

METHODS: Two scanners were used: scanner-A and scanner-B. In the phantom study, we compared the contrast enhancement between the scanners at tube voltage peaks of 80, 100 and 120 kVp. Then, we calculated the effective energies of the two CT scanners. In the clinical study, 40 patients underwent abdominal scanning with scanner-A and another 40 patients with scanner-B, with each group using the same scanning protocol. The contrast enhancement of abdominal organs was assessed quantitatively (based on the absolute difference between the attenuation of unenhanced scans and contrast-enhanced scans) and qualitatively. A two-tailed independent Student's t-test and or the Mann-Whitney U test were used to compare the discrepancies.

RESULTS: In the phantom study, contrast enhancement for scanner-B was 36.9, 32.6 and 30.8% higher than that for scanner-A at 80, 100 and 120 kVp, respectively. The effective energies were higher for scanner-A than for scanner-B. In the quantitative analysis for the clinical study, scanner-B yielded significantly better contrast enhancement of the hepatic parenchyma, pancreas, kidney, portal vein and inferior vena cava compared with that of scanner-A. The mean visual scores for contrast enhancement were also significantly higher on images obtained by scanner-B than those by scanner-A.

CONCLUSION: There were significant differences in contrast enhancement of the abdominal organs between the compared CT scanners from two different vendors even at the same scanning and contrast parameters. Advances in knowledge: Awareness of the impact of different X-ray energies on the resultant attenuation of contrast material is important when interpreting clinical CT images.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

Managing Alcohol Withdrawal Syndrome.Annals of Emergency Medicine 2024 March 26

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app