COMPARATIVE STUDY
JOURNAL ARTICLE
Add like
Add dislike
Add to saved papers

Characterization and outcomes of reinterventions in Food and Drug Administration-approved versus trial endovascular aneurysm repair devices.

OBJECTIVE: Published rates of reintervention after endovascular aneurysm repair (EVAR) range from 10% to 30%. We evaluated a single university center's experience with reinterventions in the context of Food and Drug Administration (FDA)-approved and trial devices.

METHODS: Retrospective data collection was performed for patients who underwent infrarenal EVAR and required reintervention from 2000 to 2016. Trial devices included those used in FDA feasibility and pivotal trials. Time-to-event analysis was performed using Cox regression. Predictors of mortality and explantation were evaluated using logistic regression; survival analysis was performed using Kaplan-Meier methods.

RESULTS: From 2000 to 2016, there were 1835 EVARs performed, and 137 patients (116 men; mean age, 72.2 ± 10.0 years) underwent reintervention with a mean aneurysm size of 5.9 ± 1.2 cm. The median follow-up was 5 years with an overall survival of 70.1%. The overall reintervention rate was 7.5%. FDA-approved devices had a reintervention rate of 6.4%, whereas trial devices had a rate of 14.4% (P < .001). For all devices, the most common cause of reintervention was type II endoleak (52.5%), followed by type I endoleak (18.2%), type III endoleak (9.5%), limb kink (7.3%), iliac occlusive disease (5.8%), endotension (1.5%), and other. The overall mean time to first reintervention was 2.3 ± 2.5 years, and univariate Cox regression identified male gender (hazard ratio, 1.91; 95% confidence interval [CI], 1.17-3.10; P = .010) and age at the time of EVAR (hazard ratio, 1.03; 95% CI, 1.01-1.05; P = .006) as risk factors for time to first reintervention. Among patients requiring reintervention, the mean number of reinterventions for trial devices was significantly greater than that for FDA-approved devices (2.18 vs 1.65; P = .01). Trial devices requiring reintervention had a nearly threefold higher odds for the need for more than two reinterventions (odds ratio, 2.88; 95% CI, 1.12-7.37; P = .034). Trial device, cause of reintervention, and type of reintervention were not predictive of the need for explantation or mortality, but the number of reinterventions was significantly associated with the need for explantation (odds ratio, 1.86; 95% CI, 1.17-2.96; P = .012). EVAR device and the need for explantation did not have an impact on mortality.

CONCLUSIONS: Despite the rigorous nature of patient enrollment in clinical trials and the development of newer iterations of investigational devices, patients undergoing EVAR with trial devices are more likely to undergo a greater number of reinterventions than with FDA-approved devices. Although mortality and the need for explantation were not significantly associated with trial devices, the finding of a greater number of reinterventions highlights the need to properly inform patients willing to partake in investigational device trials.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app