Transcatheter occlusion of the patent ductus arteriosus in premature infants weighing less than 1200 g.

OBJECTIVES: Over the last few decades different strategies have been proposed to treat persistent ductal patency in premature infants. The advent of the Amplatzer Duct Occluder II Additional Size (ADOIIAS) provided the potential to close the patent ductus arteriosus (PDA). Opinions differ on the significance and treatment of PDA in premature neonates. Because surgical ligation and medical therapy both have their drawbacks, interventional catheterisation can be considered as an alternative means of closing the ductus arteriosus. Our aim was to analyse the feasibility, safety and efficacy of this device in premature infants weighing <1200 g at procedure.

METHODS: Eighteen premature infants underwent transcatheter closure. The procedure was performed in the catheterisation laboratory by venous cannulation without angiography. The position of the occluder was directed by X-ray and ultrasound. We looked at procedural details, device size selection, complications and short-term and mid-term outcomes.

RESULTS: Eighteen infants born at gestational ages ranging between 23.6 and 29+6 weeks (mean±SD 25+6±3 weeks) underwent transcatheter PDA closure. Their mean age and weight at the time of the procedure was 20 days (range 8-44 days) and 980 g (range 680-1200 g), respectively. The mean PDA and device waist diameters were 3.2±0.6 mm (range 2.2-4 mm) and 4.5±0.6 mm, respectively, and the mean PDA and device lengths were 4.3±1.2 mm (range 2-10 mm) and 2.5±0.9 mm, respectively. Complete closure was achieved in all but one patient. There was no device migration. One patient developed a left pulmonary artery obstruction. Three infants died. Two deaths were related to complications of prematurity and one to the procedure.

CONCLUSIONS: Transcatheter closure of a PDA is feasible in very low weight infants with ADOIIAS and is an alternative to surgery. Success requires perfect selection and placement of the occluder.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app