We have located links that may give you full text access.
COMPARATIVE STUDY
JOURNAL ARTICLE
Effectiveness of enhanced pulse oximetry sonifications for conveying oxygen saturation ranges: a laboratory comparison of five auditory displays.
British Journal of Anaesthesia 2017 December 2
Background: Anaesthetists monitor auditory information about a patient's vital signs in an environment that can be noisy and while performing other cognitively demanding tasks. It can be difficult to identify oxygen saturation (SpO2) values using existing pulse oximeter auditory displays (sonifications).
Methods: In a laboratory setting, we compared the ability of non-clinician participants to detect transitions into and out of an SpO2 target range using five different sonifications while they performed a secondary distractor arithmetic task in the presence of background noise. The control sonification was based on the auditory display of current pulse oximeters and comprised a variable pitch with an alarm. The four experimental conditions included an Alarm Only condition, a Variable pitch only condition, and two conditions using sonifications enhanced with additional sound dimensions. Accuracy to detect SpO2 target transitions was the primary outcome.
Results: We found that participants using the two sonifications enhanced with the additional sound dimensions of tremolo and brightness were significantly more accurate (83 and 96%, respectively) at detecting transitions to and from a target SpO2 range than participants using a pitch only sonification plus alarms (57%) as implemented in current pulse oximeters.
Conclusions: Enhanced sonifications are more informative than conventional sonification. The implication is that they might allow anaesthetists to judge better when desaturation decreases below, or returns to, a target range.
Methods: In a laboratory setting, we compared the ability of non-clinician participants to detect transitions into and out of an SpO2 target range using five different sonifications while they performed a secondary distractor arithmetic task in the presence of background noise. The control sonification was based on the auditory display of current pulse oximeters and comprised a variable pitch with an alarm. The four experimental conditions included an Alarm Only condition, a Variable pitch only condition, and two conditions using sonifications enhanced with additional sound dimensions. Accuracy to detect SpO2 target transitions was the primary outcome.
Results: We found that participants using the two sonifications enhanced with the additional sound dimensions of tremolo and brightness were significantly more accurate (83 and 96%, respectively) at detecting transitions to and from a target SpO2 range than participants using a pitch only sonification plus alarms (57%) as implemented in current pulse oximeters.
Conclusions: Enhanced sonifications are more informative than conventional sonification. The implication is that they might allow anaesthetists to judge better when desaturation decreases below, or returns to, a target range.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app