Validated low-volume aldosterone immunoassay tailored to GCLP-compliant investigations in small sample volumes.
Practical Laboratory Medicine 2017 December
INTRODUCTION: Heart failure is well investigated in adults, but data in children is lacking. To overcome this shortage of reliable data, appropriate bioanalytical assays are required.
OBJECTIVES: Development and validation of a bioanalytical assay for the determination of aldosterone concentrations in small sample volumes applicable to clinical studies under Good Clinical Laboratory Practice.
METHODS: An immunoassay was developed based on a commercially available enzyme-linked immunosorbent assay and validated according to current bioanalytical guidelines of the EMA and FDA.
RESULTS: The assay (range 31.3-1000 pg/mL [86.9-2775 pmol/L]) is characterized by a between-run accuracy from - 3.8% to - 0.8% and a between-run imprecision ranging from 4.9% to 8.9% (coefficient of variation). For within-run accuracy, the relative error was between - 11.1% and + 9.0%, while within-run imprecision ranged from 1.2% to 11.8% (CV). For parallelism and dilutional linearity, the relative error of back-calculated concentrations varied from - 14.1% to + 8.4% and from - 7.4% to + 10.5%, respectively.
CONCLUSIONS: The immunoassay is compliant with the bioanalytical guidelines of the EMA and FDA and allows accurate and precise aldosterone determinations. As the assay can run low-volume samples, it is especially valuable for pediatric investigations.
OBJECTIVES: Development and validation of a bioanalytical assay for the determination of aldosterone concentrations in small sample volumes applicable to clinical studies under Good Clinical Laboratory Practice.
METHODS: An immunoassay was developed based on a commercially available enzyme-linked immunosorbent assay and validated according to current bioanalytical guidelines of the EMA and FDA.
RESULTS: The assay (range 31.3-1000 pg/mL [86.9-2775 pmol/L]) is characterized by a between-run accuracy from - 3.8% to - 0.8% and a between-run imprecision ranging from 4.9% to 8.9% (coefficient of variation). For within-run accuracy, the relative error was between - 11.1% and + 9.0%, while within-run imprecision ranged from 1.2% to 11.8% (CV). For parallelism and dilutional linearity, the relative error of back-calculated concentrations varied from - 14.1% to + 8.4% and from - 7.4% to + 10.5%, respectively.
CONCLUSIONS: The immunoassay is compliant with the bioanalytical guidelines of the EMA and FDA and allows accurate and precise aldosterone determinations. As the assay can run low-volume samples, it is especially valuable for pediatric investigations.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app