A Preclinical Evaluation Comparing the Performance of a Novel 19-G Flexible Needle to a Commercially Available 22-G EBUS-TBNA Sampling Needle.

BACKGROUND: Needle samples may not provide sufficient diagnostic material for the assessment of mediastinal lymph nodes.

OBJECTIVE: The study compared the specimen size and diagnostic performance of a new 19-G endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) needle to that of a standard 22-G EBUS-TBNA needle in a swine model of granulomatous lymphadenopathy.

METHODS: Granulomatous inflammation was induced in mediastinal lymph nodes (LNs) of 10 domestic swine by injection of talc slurry. The affected LNs were sampled with the 19- and 22-G needles. Collected core tissue area and volume were determined using a specialized software and known needle internal diameter. The sample's quality was assessed using the biopsy core morphology grade (BCMG) as well as the biopsy diagnostic correlation grade (BDCG).

RESULTS: There was a significant increase in the average LN size from baseline (11.6 ± 3.2 to 15.2 ± 3.8 mm; p < 0.001) after talc injection. A total of 132 paired samples were collected from 38 LNs. The average mass and volume of the 19-G needle sample were larger than those of the 22-G needle sample: 33.78 ± 47.48 vs. 25.18 ± 32.08 mg (p < 0.002) and 11.40 ± 13.91 vs. 6.91 ± 6.42 mm3 (p < 0.0004), respectively. The pooled needle biopsy samples for the 19- and the 22-G needles had similar BCMG (1.38 ± 0.86 vs. 1.43 ± 0.87, p > 0.2) and BDCG (1.54 ± 0.93 vs. 1.57 ± 0.93, p > 0.2). The 19-G needle samples tended towards less blood contamination (p = 0.057), more often granuloma identification (46 vs. 32%, p = 0.2) and had more cartilage contamination (0.49 ± 1.46 vs. 4.81 ± 16.49% p < 0.003).

CONCLUSION: In experienced hands, the 19- and the 22-G EBUS-TBNA needles have a similar diagnostic yield in the swine model of granulomatous lymphadenopathy. The samples collected by the 19-G needle are larger and may have less blood contamination.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app