Safety profiles of percutaneous left atrial appendage closure devices: An analysis of the Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) database from 2009 to 2016.

BACKGROUND: Percutaneous left atrial appendage closure (LAAC) is a viable option for AF patients who are unable to tolerate long-term oral anticoagulation (OAC).

OBJECTIVE: We sought to assess the safety of two commonly used percutaneous devices for LAA closure in the United States by analysis of surveillance data from the FDA Manufacturer and User Facility Device Experience (MAUDE) database.

METHODS: The MAUDE database was queried between May 1, 2006 and May 1, 2016 for LARIAT® (SentreHEART Inc., Redwood City, CA, USA) and WATCHMAN™ (Boston Scientific Corp., Marlborough, MA, USA) devices. Among 622 retrieved medical device reports, 356 unique and relevant reports were analyzed. The cumulative incidence of safety events was calculated over the study period and compared between the two devices.

RESULTS: LAAC was performed with LARIAT in 4,889 cases. WATCHMAN was implanted in 2,027 patients prior to FDA approval in March 2015 and 3,822 patients postapproval. The composite outcome of stroke/TIA, pericardiocentesis, cardiac surgery, and death occurred more frequently with WATCHMAN (cumulative incidence, 1.93% vs. 1.15%; P = 0.001). The same phenomenon was observed when comparing the WATCHMAN pre- and postapproval experiences for the composite outcome, as well as device embolization, cardiac surgery, and myocardial infarction.

CONCLUSIONS: MAUDE-reported data show that postapproval, new technology adoption is fraught with increased complications. Improved collaboration between operators, device manufacturers, and regulators can better serve patients through increased transparency and practical postmarket training and monitoring mechanisms.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app