Intradevice misalignment predicts residual leak in patients undergoing left atrial appendage closure.

AIMS: Postdeployment mutual orientation between the disk and the lobe in patients undergoing left atrial appendage closure with Amplatzer cardiac plug/Amulet device might impact on the risk of residual leak during follow-up. Thus, we evaluated in an exploratory, pilot study whether the degree of intradevice misalignment, measured by cardiac computed tomography (CT), discriminates the occurrence of peridevice leak in those patients.

METHODS: All patients (N = 15) undergoing percutaneous left atrial appendage closure with those specific devices between April 2013 and January 2015 were prospectively included. All patients received follow-up evaluation by cardiac CT at 6 months after the intervention to calculate the angle of misalignment within the device and to detect presence of residual peridevice leak.

RESULTS: The angle of misalignment between the disk and the lobe of the device significantly discriminated between patients with and without peridevice leak (area under the curve 0.96, 95% confidence interval 0.88-1.0; P = 0.003), with an angle more than 20° being associated with nine-fold higher risk of residual leak. This angle of intradevice misalignment calculated by cardiac CT was significantly correlated with that measured by two-dimensional transesophageal echocardiography or X-rays in the cath lab after the device deployment (r = 0.943 and r = 0.938, respectively).

CONCLUSION: A marked intradevice misalignment after Amplatzer cardiac plug/Amulet device implantation significantly predicts the occurrence of postprocedural leak in patients with atrial fibrillation; if confirmed by larger studies, these findings may impact on practice patterns.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app