Reliability and Reproducibility of Subject Positioning with EOS Low-Dose Biplanar X-ray.
BACKGROUND: EOS low-dose biplanar X-ray used with tantalum bead implantation is an appealing imaging modality to evaluate limb length and physeal growth due its relatively low radiation exposure, excellent inter- and intra-observer reliability, and minimal magnification/shrinkage error.
QUESTIONS/PURPOSES: The purpose of this study was to establish the error in total length and inter-bead distances using EOS and tantalum beads due to variable positioning in the EOS gantry, by assessing variation in measurements made on the same subject repeatedly positioning by one positioner (intra-positioner measurement error) and variation in measurements made on the same subject with positioning by multiple positioners (inter-positioner measurement error).
METHODS: Tantalum bead markers were placed about the distal femoral physis of a cadaveric lamb femur. Three investigators positioned the femur in the EOS gantry 10 times, totaling 30 EOS scans. Total limb length and inter-bead distances were measured on AP and lateral views; mean and standard error were calculated. A random effects analysis of variance for nested data was used to determine the proportion of variation due to differences between positioners and differences between positioning by the same positioner.
RESULTS: Intra-positioner measurement error ranged from 0.01 to 0.06 mm. Inter-positioner measurement error ranged from 0.00 to 0.09 mm.
CONCLUSIONS: EOS has relatively low radiation and allows standing assessment of limb length and alignment. In this study, length measurements and inter-bead distances demonstrated minimal error due to positioning in the EOS gantry, documenting that there is minimal error from positioning, minimal magnification/shrinkage error, and exceptional inter and intra-rater reliability. EOS is the preferred method for length and angular measurements, and with tantalum beads, is ideal for measuring growth about the physis.
QUESTIONS/PURPOSES: The purpose of this study was to establish the error in total length and inter-bead distances using EOS and tantalum beads due to variable positioning in the EOS gantry, by assessing variation in measurements made on the same subject repeatedly positioning by one positioner (intra-positioner measurement error) and variation in measurements made on the same subject with positioning by multiple positioners (inter-positioner measurement error).
METHODS: Tantalum bead markers were placed about the distal femoral physis of a cadaveric lamb femur. Three investigators positioned the femur in the EOS gantry 10 times, totaling 30 EOS scans. Total limb length and inter-bead distances were measured on AP and lateral views; mean and standard error were calculated. A random effects analysis of variance for nested data was used to determine the proportion of variation due to differences between positioners and differences between positioning by the same positioner.
RESULTS: Intra-positioner measurement error ranged from 0.01 to 0.06 mm. Inter-positioner measurement error ranged from 0.00 to 0.09 mm.
CONCLUSIONS: EOS has relatively low radiation and allows standing assessment of limb length and alignment. In this study, length measurements and inter-bead distances demonstrated minimal error due to positioning in the EOS gantry, documenting that there is minimal error from positioning, minimal magnification/shrinkage error, and exceptional inter and intra-rater reliability. EOS is the preferred method for length and angular measurements, and with tantalum beads, is ideal for measuring growth about the physis.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app