It is time for a better clip applier - 3 mm, percutaneous, non-crushing and locking.
INTRODUCTION: Since the advent of laparoscopy there have been attempts to minimize abdominal wall incisions. For this purpose smaller instruments have been produced. Our aim was to develop the first 3 mm percutaneous clip applier and to make it better than the standard clips of today.
MATERIAL AND METHODS: The ClipTip clip is made of Nitinol and has a crocodile shaped jaws, which when apposed effectively seal vessels. The shaft operates as a retractable needle permitting percutaneous insertion. Closing, reopening and reclosing is possible. The physical properties of the device were compared to three commercially available clip appliers. Surgeries were performed on porcine animals by experienced surgeons.
RESULTS: In comparison to available clips, the superiority of the ClipTip is a combination of wide effective length alongside the ability to withstand strong forces. In live animal studies the Cliptip was inserted into the peritoneal cavity without any injuries. Vessels were ligated successfully and no clip dislodgement or leakage occurred.
CONCLUSIONS: We developed the next generation clip applier with better properties. Advantages include its length, the needleoscopic caliber, non-crushing effect, locking mechanism and wide aperture. The device has performed safely and effectively in pre-clinical tests. Further studies are planned in humans.
MATERIAL AND METHODS: The ClipTip clip is made of Nitinol and has a crocodile shaped jaws, which when apposed effectively seal vessels. The shaft operates as a retractable needle permitting percutaneous insertion. Closing, reopening and reclosing is possible. The physical properties of the device were compared to three commercially available clip appliers. Surgeries were performed on porcine animals by experienced surgeons.
RESULTS: In comparison to available clips, the superiority of the ClipTip is a combination of wide effective length alongside the ability to withstand strong forces. In live animal studies the Cliptip was inserted into the peritoneal cavity without any injuries. Vessels were ligated successfully and no clip dislodgement or leakage occurred.
CONCLUSIONS: We developed the next generation clip applier with better properties. Advantages include its length, the needleoscopic caliber, non-crushing effect, locking mechanism and wide aperture. The device has performed safely and effectively in pre-clinical tests. Further studies are planned in humans.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
The Effect of Albumin Administration in Critically Ill Patients: A Retrospective Single-Center Analysis.Critical Care Medicine 2024 Februrary 8
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app