Add like
Add dislike
Add to saved papers

Cost-effectiveness of a risk-stratified approach to cardiac resynchronisation therapy defibrillators (high versus low) at the time of generator change.

Heart 2018 March
OBJECTIVE: Responders to cardiac resynchronisation therapy whose device has a defibrillator component and who do not receive a therapy in the lifetime of the first generator have a very low incidence of appropriate therapy after box change. We investigated the cost implications of using a risk stratification tool at the time of generator change resulting in these patients being reimplanted with a resynchronisation pacemaker.

METHODS: A decision tree was created using previously published data which had demonstrated an annualised appropriate defibrillator therapy risk of 2.33%. Costs were calculated at National Health Service (NHS) national tariff rates (2016-2017). EQ-5D utility values were applied to device reimplantations, admissions and mortality data, which were then used to estimate quality-adjusted life-years (QALYs) over 5 years.

RESULTS: At 5 years, the incremental cost of replacing a resynchronisation defibrillator device with a second resynchronisation defibrillator versus resynchronisation pacemaker was £5045 per patient. Incremental QALY gained was 0.0165 (defibrillator vs pacemaker), resulting in an incremental cost-effectiveness ratio (ICER) of £305 712 per QALYs gained. Probabilistic sensitivity analysis resulted in an ICER of £313 612 (defibrillator vs pacemaker). For reimplantation of all patients with a defibrillator rather than a pacemaker to yield an ICER of less than £30 000 per QALY gained (current NHS cut-off for approval of treatment), the annual arrhythmic event rate would need to be 9.3%. The budget impact of selective replacement was a saving of £2 133 985 per year.

CONCLUSIONS: Implanting low-risk patients with a resynchronisation defibrillator with the same device at the time of generator change is not cost-effective by current NHS criteria. Further research is required to understand the impact of these findings on individual patients at the time of generator change.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app