We have located links that may give you full text access.
Assessment of a photon-counting detector for a dual-energy C-arm angiographic system.
Medical Physics 2017 November
PURPOSE: This article presents the implementation and assessment of photon-counting dual-energy x-ray detector technology for angiographic C-arm systems in interventional radiology.
METHODS: A photon-counting detector was successfully integrated into a clinical C-arm CT system. Detector performance was assessed using image uniformity metrics in both 2D projections and 3D cone-beam computed tomography (CBCT) images. Uniform exposure fields were acquired to analyze projection images and scans of a homogeneous cylinder phantom were taken to analyze 3D reconstructions. Image uniformity was assessed over a broad range of imaging parameters.
RESULTS: Detector calibration greatly improved image uniformity, reducing image variation from 8.8% to 0.5% in an ideal scenario, but image uniformity degraded when imaging parameters varied strongly from values set at calibration: the tube voltage, low-high energy threshhold, and tube current had the greatest impact. Material discrimination and dynamic angiography capabilities were successfully demonstrated in separate phantom and in vivo experiments.
CONCLUSION: The uniformity results identified major factors degrading image quality. The quantitative results will guide selection of calibration points to mitigate the loss of uniformity. The unique combination of dual-energy and fluoroscopy imaging capabilities with a flat-panel photon-counting detector may enable new applications in interventional radiology.
METHODS: A photon-counting detector was successfully integrated into a clinical C-arm CT system. Detector performance was assessed using image uniformity metrics in both 2D projections and 3D cone-beam computed tomography (CBCT) images. Uniform exposure fields were acquired to analyze projection images and scans of a homogeneous cylinder phantom were taken to analyze 3D reconstructions. Image uniformity was assessed over a broad range of imaging parameters.
RESULTS: Detector calibration greatly improved image uniformity, reducing image variation from 8.8% to 0.5% in an ideal scenario, but image uniformity degraded when imaging parameters varied strongly from values set at calibration: the tube voltage, low-high energy threshhold, and tube current had the greatest impact. Material discrimination and dynamic angiography capabilities were successfully demonstrated in separate phantom and in vivo experiments.
CONCLUSION: The uniformity results identified major factors degrading image quality. The quantitative results will guide selection of calibration points to mitigate the loss of uniformity. The unique combination of dual-energy and fluoroscopy imaging capabilities with a flat-panel photon-counting detector may enable new applications in interventional radiology.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app